![KaVo SMARTtorque S609 C Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/kavo/smarttorque-s609-c/smarttorque-s609-c_instructions-for-use-manual_3987786006.webp)
Instructions for use SMARTtorque S609 C - 1.008.1644 SMARTtorque Mini S605 C - 1.008.1645 SMARTtorque
S608 C - 3.001.6560
2 Safety | 2.1 Infection hazard
6 / 26
2 Safety
The instructions for use are a component of the product and must be read care-
fully prior to use and be accessible at all times.
The device may only be used in accordance with the intended use, any other
type of use is not permitted.
2.1 Infection hazard
Patients, users or third parties can be infected by contaminated medical de-
vices.
▶
Take suitable personal protective measures.
▶
Follow the instructions for use of the components.
▶
Before initial startup and after each use, process the product and acces-
sories appropriately.
▶
Carry out the processing as described in the instructions for use. The proce-
dure has been validated by the manufacturer.
▶
If you deviate from this procedure, it is essential to make sure that the pro-
cessing is effective.
▶
Process the product and accessories appropriately before disposal.
▶
In the case of injury to soft tissue, do not continue treatment in the oral
cavity with instruments driven by compressed air.
2.2 Technical condition
A damaged device or components can injure patients, users and third parties.
▶
Use the device and components only if there is no damage on the outside.
▶
Check to make sure that the device is working properly and is in satisfac-
tory condition before each use.
▶
Have parts with sites of breakage or surface changes checked by the Ser-
vice.
▶
If the following defects occur, stop working and have the service personnel
carry out repair work:
▪ Malfunctions
▪ Damage
▪ Irregular running noise
▪ Excessive vibration
▪ Overheating
▪ Dental bur or diamond is not firmly locked in the handpiece
To ensure optimum function and to prevent property damage, please comply
with the following instructions:
▶
Service the medical device with care products and systems regularly as de-
scribed in the instructions for use.
▶
The product should be processed and stored in a dry location, according to
instructions, if it is not to be used for an extended period of time.