![KaVo SMARTtorque S609 C Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/kavo/smarttorque-s609-c/smarttorque-s609-c_instructions-for-use-manual_3987786013.webp)
Instructions for use SMARTtorque S609 C - 1.008.1644 SMARTtorque Mini S605 C - 1.008.1645 SMARTtorque
S608 C - 3.001.6560
5 Operation | 5.1 Attaching the medical device
5 Operation
Note
At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20-30 seconds after each patient.
5.1 Attaching the medical device
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked can release from the hose during
treatment.
▶
Before each use, check if the medical device is securely locked onto the
hose by pulling on it.
▶
Connect the coupling part of the medical device to the hose proceeding in
clockwise direction.
▶
Pull on it to make sure that the medical device is securely attached to the
hose.
Note
The water quantity can be regulated through the coupling or the treatment
centre.
5.2 Removing the medical device
▶
Firmly grasp the coupling and pull the medical device off while twisting
slightly.
5.3 Inserting the bur or diamond
Note
Only use carbide burs or diamonds that comply with ISO 1797-1 type 3, are
made of steel or hard metal and meet the following criteria:
- Shaft diameter: 1.59 to 1.60 mm
- Blade diameter: max. 2 mm
SMARTtorque S608 C and SMARTtorque S609 C:
- Overall length: max. 25 mm
- Shaft clamping length: min. 11 mm
SMARTtorque Mini S605 C:
- Overall length: max. 21 mm
- Shaft clamping length: min. 9 mm
WARNING
Use recommended dental burs or diamonds only.
Risk of injury.
▶
Comply with the instructions for use and the intended use of the bur.
13 / 26
Содержание SMARTtorque S609 C
Страница 25: ......
Страница 26: ...1 008 1914 bd 20190813 06 en US...