5 Operation | 5.4 Removing the bur or diamond

14 / 26

 CAUTION

Use of burs with worn or damaged shafts.

Risk of injury, tool may fall out during treatment.

Never use burs with damaged or worn shafts.

 CAUTION

Danger of injury from bur.

Infections or cuts.

Wear gloves or finger guards.

 CAUTION

Hazard from defective chucking system.

The bur could fall out and cause injury.

Pull on the bur to check if the clamping system works properly and the tool
is firmly clamped. Wear gloves or a finger guard when you check, insert or
remove the bits to prevent injury and infection.

NOTICE

Tool shaft slips inside the chuck due to excessive speed of the tool or
abrupt engagement of the tool.

Property damage to tool shaft and chuck system, reduction of the service life of
tool and chuck system.

Do not operate the tool at a higher speed than recommended by the manu-
facturer.

NOTICE

Use of burs with worn or damaged shafts.

Property damage to the chuck system, tool is difficult or impossible to remove
from the chuck system.

Do not use burs with damaged or worn shafts.

Forcefully press the push-button with your thumb and simultaneously insert
the bur to the limit stop.

Check that the bur is securely attached by pulling on it.

5.4 Removing the bur or diamond

 WARNING

Hazard from rotating bur.

Cuts, infection and burn injury.

Never press the push-button while the bur is rotating.

Do not touch the bur while it is rotating!

Remove the bur from the contra-angle handpiece after treatment to avoid
injury and infection when putting it away.

NOTICE

Damage to the chucking system.

Property damage

Do not press the push-button while the bur is rotating.

Содержание SMARTtorque S609 C

Страница 1: ...Instructions for use SMARTtorque S609 C 1 008 1644 SMARTtorque Mini S605 C 1 008 1645 SMARTtorque S608 C 3 001 6560...

Страница 2: ...stributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo c...

Страница 3: ...oting 16 6 1 Cleaning the spray nozzle 16 7 Processing steps in accordance with ISO 17664 17 7 1 Preparations at the site of use 17 7 2 Manual processing 17 7 2 1 Manual external cleaning 17 7 2 2 Man...

Страница 4: ...airs please contact KaVo Repair Service For scheduling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll fre...

Страница 5: ...tion and storage conditions Air pressure Transportation and storage conditions Humidity range Protect from moisture Keep dry Fragile handle with care HIBC Code Hazard levels The warning and safety not...

Страница 6: ...ct and accessories appropriately before disposal In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 2 2 Technical condition A d...

Страница 7: ...to the device unless these have been ap proved by the manufacturer of the product Use original KaVo spare parts only 2 4 Qualification of personnel Application of the product by users lacking appropr...

Страница 8: ...ovided with a fiber optic light system The devices can be sterilized in a steam sterilizer autoclave Through the tube and the MULTIflex coupling con nected to a dental unit the SMARTtorque handpieces...

Страница 9: ...C Water consumption 50 ml min Idle speed 350 000 to 400 000 min 1 350 000 to 450 000 min 1 350 000 to 400 000 min 1 Power torque 16 W 0 14 Ncm 13 W 0 12 Ncm 16 W 0 14 Ncm Head size Default Mini Defaul...

Страница 10: ...e Mini S605 C 1 008 1645 SMARTtorque S608 C 3 001 6560 3 Product description 3 3 Transportation and storage conditions 10 26 Relative humidity 5 RH to 95 RH absence of condensation Air pressure 700 hP...

Страница 11: ...process and sterilize the product and accessories appropri ately See also 2 7 Processing steps in accordance with ISO 17664 Page 17 4 1 Checking the water quantity CAUTION Overheating of the tooth due...

Страница 12: ...gauge Mat no 0 411 8731 between hose and medical device and check the following pressures Drive pressure drive pressure recommended return air pressure spray water pressure and spray air pressure SMA...

Страница 13: ...ling on it Connect the coupling part of the medical device to the hose proceeding in clockwise direction Pull on it to make sure that the medical device is securely attached to the hose Note The water...

Страница 14: ...ement of the tool Property damage to tool shaft and chuck system reduction of the service life of tool and chuck system Do not operate the tool at a higher speed than recommended by the manu facturer...

Страница 15: ...008 1644 SMARTtorque Mini S605 C 1 008 1645 SMARTtorque S608 C 3 001 6560 5 Operation 5 4 Removing the bur or diamond After the bur has stopped rotating firmly press the push button with your thumb a...

Страница 16: ...ing the spray nozzle WARNING Hazard from non sterile products Infection hazard for dentist and patient Reprocess and sterilize the medical device properly before the next use CAUTION Hazard from insuf...

Страница 17: ...tool Remove the tool Process the medical device as soon as possible after treatment To minimize the risk of infection during processing always wear protective gloves Remove the tool from the medical...

Страница 18: ...d by the disinfectant manufacturer and proven by an expert opinion Approved disinfectants CaviWipes and CaviCide made by Metrex Consumables required Cloths for wiping the medical device Spray the disi...

Страница 19: ...nds washer disinfectors in accordance with ISO 15883 1 which are operated using alkaline cleaning agents For program settings as well as cleansers and disinfectants to be used please refer to the Inst...

Страница 20: ...ly Remove the tool from the medical device Position the tip of the spray nipple in the opening and apply the spray Press the spray key once for 1 2 seconds 7 4 2 Servicing with KaVo QUATTROcare PLUS S...

Страница 21: ...e minutes with no service procedure running the device automatically switches back to normal service mode See also 2 Servicing with KaVo QUATTROcare PLUS 7 5 Packaging Note The sterile goods package m...

Страница 22: ...135 C 275 F Drying time 16 min Sterilizer using the gravity method at least 10 minutes at 135 C 275 F drying time 30 minutes or alternatively at least 60 minutes at 121 C 250 F Drying time 15 min Remo...

Страница 23: ...o Spare turbine Standard 1 008 4163 Spare turbine Mini 1 008 4164 Wrench Standard 0 411 3053 Wrench Mini new 1 008 6133 Wrench Mini old 1 006 3384 Nozzle pin 0 410 0921 INTRA spray head incl Midwest a...

Страница 24: ...contrary KaVo shall not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenanc...

Страница 25: ......

Страница 26: ...1 008 1914 bd 20190813 06 en US...

Отзывы: