5 Troubleshooting
5.1 Tips for operators and service technicians
NOTE
Any repairs exceeding routine maintenance may only be carried out by
qualified personnel or our service.
Prior to working on the device or in case of danger, disconnect it from the
mains (e. g. pull the mains plug).
5.1.1 Poor X-ray image
Instead of the X-ray image, software shows completely white image or no
image
Possible cause
Remedy
Imaging plate not fed in straight and inactive side
scanned
Scan the imaging plate again immediately,
protecting it against ambient light and making
sure you feed it in correctly in the process.
Image data on the imaging plate has been
erased, e.g. by ambient light
Always scan the image data of the imaging plate
as quickly as possible.
Fault on the unit
Inform a Service Technician.
No image data on imaging plate, imaging plate
not exposed or not sufficiently exposed
X-ray tubes / check unit settings
Expose the imaging plate.
X-ray unit is faulty
Inform a Service Technician.
Incorrect cartridge, light protection cover was
also pushed into the unit
Use the correct cartridge for the size of imaging
plate being used.
Imaging plate falls out of unit and no image appears on monitor
Possible cause
Remedy
Correct imaging plate not used
Only use distributor's imaging plates.
X-ray image too dark
Possible cause
Remedy
X-ray dose too high
Check X-ray parameters.
Incorrect brightness/contrast settings in the
software
Adjust the brightness of the X-ray image in the
software.
X-ray image too bright
Possible cause
Remedy
Exposed imaging plate has been exposed to
ambient light
Always scan the image data of the imaging plate
as quickly as possible.
X-ray dose too low
Check X-ray parameters.
Incorrect brightness/contrast settings in the
software
Adjust the brightness of the X-ray image in the
software.
5 Troubleshooting
52 KaVo ProXam iP
Instructions for use