3. Disinfect the light protection cover and protective gloves with a suitable
disinfection wipe (see section "Light protection cover" on page 50).
4. Allow the light protection cover to fully dry.
5. Pull off the protective gloves and disinfect the hands.
CAUTION
Powder from the protective gloves on the imaging plate can damage the
unit during scanning
Completely clean all traces of the protective glove powder from your
hands before handling the imaging plate.
6. Tear off the light protection cover.
4.2.3 Scanning image data
4.2.3.1 Starting imaging plate scanner and software
About this task
NOTE
The reading-out process is described using the Romexis imaging software.
For further information see the
Romexis 6 User's manual, section Capturing
intraoral images with imaging plate scanner.
Steps
1. Press the on / off switch to switch on the device.
2. Switch on the computer and monitor.
3. Start Romexis.
4. Select the patient.
5. Select the PSP capturing button.
Recording starts directly. The green status LED lights up.
6. Scan the imaging plate at this point (and not before).
4 Usage
Instructions for use
KaVo ProXam iP 43