4.3.2 Light protection cover
The surface of the unit must be cleaned and disinfected if it is contaminated
or visibly soiled.
•
Disinfect the light protection cover using a disinfectant before and
after placement; the manufacturer recommends FD 333 forte wipes
(virucidal), FD 350 (limited virucidal activity) and FD 322 premium wipes
(limited virucidal activity)
•
Allow the light protection cover to completely dry before using it
4.3.3 Imaging plate
Cleaning and disinfection wipes are unsuitable for cleaning imaging plates
and may cause damage to them.
Only use a cleaning agent that is compatible with the materials: the
manufacturer recommends the imaging plate cleaning wipe (see section
"Consumables" on page 9). Only this product has been subjected to material
compatibility testing.
NOTE
Heat or humidity will damage the imaging plate.
Do not steam sterilise the imaging plate.
Do not immersion-disinfect the imaging plate.
Only use cleaning agents that are compatible with the materials.
•
Soiling on both sides of the imaging plate should be cleaned off with a
soft, lint-free wipe prior to every use
•
Remove resistant or dried on dirt with the imaging plate cleaning wipe
When doing this, observe the instructions for use for the cleaning wipe.
•
Allow the imaging plate to completely dry before using it
4.3.4 Storage box with imaging plate storage tray
Clean and disinfect the surface of the storage box and the internal imaging
plate storage tray in the event of contamination or visible soiling.
The manufacturer recommends the following disinfectant for the storage box:
•
FD 366 sensitive
The manufacturer recommends the following disinfectants for the imaging
plate storage tray:
•
FD 350
•
FD 366 sensitive
Observe the following cleaning and disinfection practices:
•
Clean the surface of the storage box and the imaging plate storage tray
with a soft, lint-free cloth that has been dampened with cold tap water
•
Disinfect the storage box using a disinfection wipe; alternatively, use
disinfectant on a soft, lint-free cloth
•
Disinfect the imaging plate storage tray using a disinfection wipe
•
Alternatively, the imaging plate storage tray can be treated in a thermal
disinfector or steam steriliser
Do not exceed a temperature of 134°C when doing this.
4 Usage
50 KaVo ProXam iP
Instructions for use
Содержание ProXam iP
Страница 1: ...KaVo ProXam iP Instructions for use EN ...
Страница 67: ...A 4 Sample handover protocol template Appendix A Resolution data Instructions for use KaVo ProXam iP 63 ...
Страница 69: ......
Страница 70: ...KaVo Dental GmbH Bismarckring 39 88400 Biberach Deutschland www kavo com 1 015 1209 2160400008L02 ...