Instructions for use PROPHYflex 4
6 Troubleshooting | 6.1 Cleaning a blocked cannula
6 Troubleshooting
Preventive measures
▶
After each treatment and before each sterilisation, unscrew the powder
container in an anticlockwise direction and replace it with a clean powder
container.
▶
Mount the PROPHYflex on the MULTIflex coupling and blow through the air
and water channels.
▶
Switch the water off and blow through the air and water channels again.
6.1 Cleaning a blocked cannula
▶
Unscrew the cannula in anticlockwise direction using the wrench.
▶
Slide the nozzle needle into the cannula from the front while rotating it.
▶
Then slide the nozzle needle into the cannula from the back while rotating
it.
▶
Then remove the nozzle needle, and blow out the cannula with compressed
air.
19 / 34