Instructions for use PROPHYflex 4
6 Troubleshooting | 6.2 Cleaning a clogged main body
20 / 34
6.2 Cleaning a clogged main body
▶
Pulling off the grip sleeve with cannula.
▶
Unscrew the powder container anticlockwise.
▶
Push the nozzle needle through the aperture of the nozzle.
▶
Use the cleaning drill to clean or remove obstructions from the media tube.
▶
Then blow through with compressed air.
If the clogging persists:
▶
Use the wrench for cannula to unscrew the nozzle by placing the lateral re-
cess of the wrench on the nozzle and unscrewing the nozzle.