Instructions for use PROPHYflex 4 Perio Kit - 1.011.9403 Refill PROPHYflex perio tip - 1.010.0287
6 Reprocessing steps in accordance with ISO 17664 | 6.3 Manual Reprocessing
16 / 24
▶
If the PROPHYflex perio tip cannot be determined to be patent after the
manual rinsing procedure or once 10 reprocessing cycles have been com-
pleted, the medical device must be replaced.
6.3 Manual Reprocessing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
Clean it in a washer disinfector only.
6.3.1 Manual external cleaning
Manual external cleaning of the PROPHYflex 4 perio tip is not applicable. For ef-
fective reprocessing after non-fixating pre-cleaning, the inside of the device
must be subjected to automated cleaning in a cleaning and disinfection unit in
accordance with ISO 15883-1.
6.3.2 Manual internal cleaning
Manual internal cleaning of the PROPHYflex 4 perio tip is not applicable. For ef-
fective reprocessing after non-fixating pre-cleaning, the inside of the device
must be subjected to automated cleaning in a cleaning and disinfection unit in
accordance with ISO 15883-1.
6.3.3 Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
NOTICE
Never disinfect the medical device with chloride-containing products.
Malfunction and material damage.
▶
Only disinfect in a washer disinfector or manually.