Instructions for use PROPHYflex 4 Perio Kit - 1.011.9403 Refill PROPHYflex perio tip - 1.010.0287

6 Reprocessing steps in accordance with ISO 17664 | 6.8 Sterilisation

20 / 24

6.8 Sterilisation

Sterilisation in a steam steriliser (autoclave) according
to EN 13060 / ISO 17665-1

NOTICE

Contact corrosion due to moisture.

Damage to product.

Immediately remove the product from the steam steriliser after the steril-
isation cycle.

The KaVo medical device has a maximum temperature resistance up to 138 

(280.4 °F).

Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilisation processes:

▪ Steriliser with triple pre-vacuum:

- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)

▪ Steriliser using the gravity method:

- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)

Use according to the manufacturer's Instructions for Use.

6.9 Storage

Reprocessed products must be stored appropriately such that they are protec-
ted from germs (as far as possible) and dust, in a dry, dark, cool room.

Note

Comply with the expiry date of the sterilised items.

Содержание PROPHYflex 4 Perio Kit

Страница 1: ...Instructions for use PROPHYflex 4 Perio Kit 1 011 9403 Refill PROPHYflex perio tip 1 010 0287...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...6 1 Preparations at the site of use 15 6 2 Non fixing pre cleaning of the PROPHYflex perio tip 15 6 3 Manual Reprocessing 16 6 3 1 Manual external cleaning 16 6 3 2 Manual internal cleaning 16 6 3 3...

Страница 4: ...e contact your local dealer or the KaVo Repair Service directly 49 0 7351 56 1900 service reparatur kavokerr com Target group This document is intended for dentists and their assistants The startup se...

Страница 5: ...tation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and m...

Страница 6: ...Avoid the insufflation of spray in open wounds The improper use of the product might lead to emphysema Emphysema may arise in extreme individual cases especially in the presence of pathological gin g...

Страница 7: ...vice unless these have been ap proved by the manufacturer of the product 2 5 Qualification of personnel Application of the product by users without the appropriate medical training could injure the pa...

Страница 8: ...val 20g Perio Powder PROPHYflex perio tip Refill REF 1 010 0287 3 2 1 Index 10x perio tip 1x Wrench 1x Nozzle pin The markings of the PROPHYflex perio tip are based on the WHO index 3 1 Intended use I...

Страница 9: ...ding to these regulations the user is required to only use properly operating equipment to comply with the specified intended use to protect himself the patient and third parties from danger to avoid...

Страница 10: ...accessories appro priately Dispose of the perio tip like injection needles in a closed container See also 2 6 Reprocessing steps in accordance with ISO 17664 Page 15 4 1 Inserting the adaptor into th...

Страница 11: ...tip before each treatment and check if it is seated firmly Before each treatment make sure that the perio tip works properly Note In order to screw it on hold the wrench with 2 fingers in the round g...

Страница 12: ...x perio tip 1 010 0287 4 Startup and shut down 4 4 Removing the PROPHYflex perio tip 12 24 Carefully pull the PROPHYflex perio tip out of the wrench Take the adapter off Unscrew the adapter in counter...

Страница 13: ...w the container filled with the PROPHYflex Perio Powder onto the PRO PHYflex 4 Settings on the powder jet unit Aim the jet nozzle at a wet washbasin from a distance of approx 20 cm Set the quantities...

Страница 14: ...raindication Apical focus Root caries Treatment procedure 1 Introduce the perio tip to the floor of the pocket then pull the perio tip out by 1 2 mm 2 Start up the PROPHYflex with the foot control 3 U...

Страница 15: ...ng water any longer and strong corrosion oc curs Use the wrench to remove the PROPHYflex perio tip from the adaptor See also 2 4 4 Removing the PROPHYflex perio tip Page 11 Note Hardened or damaged ad...

Страница 16: ...ter non fixating pre cleaning the inside of the device must be subjected to automated cleaning in a cleaning and disinfection unit in accordance with ISO 15883 1 6 3 2 Manual internal cleaning Manual...

Страница 17: ...ernal disinfection of the PROPHYflex 4 perio tip is not applicable For effective reprocessing the inside of the device must be subjected to automated cleaning in a cleaning and disinfection unit in ac...

Страница 18: ...rial damage Only disinfect in a washer disinfector or manually NOTICE Never reprocess this medical device in an ultrasonic device Malfunction and material damage Clean manually or in a washer disinfec...

Страница 19: ...is normally part of the cleaning program of the thermod isinfector Note Please observe the instructions for use of the thermodisinfector In order to prevent impairment of the KaVo medical device make...

Страница 20: ...dical device has a maximum temperature resistance up to 138 280 4 F Select a suitable procedure depending on the available autoclave from the fol lowing sterilisation processes Steriliser with triple...

Страница 21: ...rom dental suppliers Adapter perio tip 1 011 9400 Key 2 000 0901 Wrench for cannula 3 004 6351 Nozzle pin 0 410 0911 PROPHYflex perio tip Refill 10 pcs 1 010 0287 Cleaning adapter PROPHYflex 4 long 3...

Страница 22: ...rary KaVo shall not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or...

Страница 23: ......

Страница 24: ...1 013 3754 bd 20180209 01 en...

Отзывы: