6 Reprocessing methods according to ISO 17664 | 6.4 Sterilisation
6.3.2 Automatic Drying
Not applicable.
6.4 Sterilisation
Note
The fibre-optic cable can be subjected to sterilisation.
Sterilization in a steam sterilizer in compliance with DIN
EN 13060
NOTICE
Damage caused by over-heating.
Damage to the fibre-optic cable.
▶
Immediately remove the parts from the steriliser after the sterilisation cycle
is completed.
▶
Weld the fibre-optic cable in sterile goods packaging.
▶
Sterilise with a triple fractionated vacuum (B class steriliser); hold for 4
minutes; 134
°
C ± 1 °C; overpressure: 2.13 bar.
6.5 Storage
▶
Reprocessed products should be stored protected from dust with minimum
exposure to germs in a dry, dark and cool place.