Instructions for use KaVo polymerisation light Poly One
6 Reprocessing methods according to ISO 17664 | 6.2 Disinfection
20 / 26
▶
Immediately remove contamination from materials used in the dental office
using a soft cloth and alcohol.
▶
Clean the handpiece with a soft, moist cloth.
6.1.2 Automated cleaning
NOTICE
Property damage caused by improper cleaning/disinfection.
▶
Do not sterilise the fibre-optic cable in a hot-air steriliser.
▶
Subject the fibre-optic cable to thermodisinfection and sterilisation.
6.2 Disinfection
6.2.1 Manual disinfection
External disinfection
Approved disinfectants:
▪ Mikrozid AF Liquid (made by Schülke & Mayr)
▪ FD 322 (Dürr)
▪ Incidin Liquid (Ecolab)
▪ CaviCide / CaviWipes (Metrex)
▶
Perform wipe disinfection only, the use of spray disinfection and disinfection
baths is not feasible.
▶
Wipe the handpiece, protective shield, fibre-optic cable, and silicone holder
with a soft cloth and approved disinfectants.
▶
Use disinfectants according to the manufacturer's Instructions for Use.
▶
Please note the disinfectant safety data sheet.
6.2.2 Automated disinfection
NOTICE
Property damage caused by improper cleaning/disinfection.
▶
Do not sterilise the fibre-optic cable in a hot-air steriliser.
▶
Subject the fibre-optic cable to thermodisinfection and sterilisation.
6.3 Drying
6.3.1 Manual drying
▶
Blow off the outside of handpiece, protective shield, fibre-optic cable, and
silicone holder with compressed air until no water drops are visible any
longer.