Instructions for use KaVo polymerisation light Poly One
6 Reprocessing methods according to ISO 17664 | 6.1 Cleaning
6 Reprocessing methods according to ISO 17664
KaVo recommends to use original KaVo parts
®
for operation and repair only,
since their safety, function, and specific suitability have been tested in extens-
ive tests.
NOTICE
Property damage caused by improper cleaning/disinfection.
▶
Do not wash the curing light Poly One in the washer disinfector.
▶
Do not sterilise the curing light Poly One in the hot-air steriliser.
▶
Do not place the curing light Poly One in a disinfection or ultrasonic bath.
Note
For cleaning/reprocessing of the fibre-optic cable of the curing light Poly One,
the fibre-optic cable must be removed from the handpiece.
6.1 Cleaning
6.1.1 Manual cleaning
Note
Combine manual cleaning with disinfection.
Internal cleaning
Not applicable.
External cleaning
Accessories required:
▪ Tap water 30
o
C ± 5
o
C
▪ Brush, e.g. medium-hard toothbrush
▶
Hold the curing light Poly One by the handpiece and disassemble the fibre-
optic cable.
▶
Disassemble the silicone holder and protective shield.
▶
Clean the fibre-optic cable, protective shield, and silicone holder with a
brush and running water (at least tap water quality).
19 / 26