Special procedures
8.1 General
When updating the unit, e.g. with a cephalometric or 3D
option, it is always recommended to update the unit with
latest firmware and the image capturing PC with the latest
driver software (KaVo Driver Update CD). Please contact
the manufacturer’s Technical Support for further
information.
NOTICE! It is compulsory to update the earlier units with a
cooling kit for the CPU before making firmware updates
(R1.05 or newer).
8.2 Cephalometric upgrade kit
For the installation of the cephalometric upgrade kit see
chapter 5.9 Installing the cephalometric unit.
For changing the handedness of the unit see chapter 8.3
Changing the cephalostat arm side.
8.3 Changing the cephalostat arm side
Left hand cephalostat arm is on the left when standing in
front of the unit. In these instructions left handed
cephalostat is changed to be right handed.
1.
Remove the upper cover (A) from the cephalostat
head.
A
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...