7 Troubleshooting
7.2 Image appearance
Problem
Possible cause
Solution
Images are too light
1.
CLINIVIEW: Contrast
and brightness not
optimum
2.
CLINIVIEW: Gamma
not set correctly
1.
Adjust contrast and
brightness.
2.
Select a more fitting
histogram type and
check gamma setting.
Images are too dark
1.
CLINIVIEW: Contrast
and brightness not
optimum.
2.
Manual technical
factors used too high.
1.
Adjust contrast and
density.
2.
Decrease technical
factors.
Lack of image contrast
1.
CLINIVIEW: Contrast
and brightness not
optimum.
2.
kV used is too high.
3.
Gamma value is not
correct for the monitor
being used.
1.
Adjust contrast and
brightness.
2.
Lower the kV setting.
3.
Adjust Gamma value
ORTHOPANTOMOGRAPH
™
OP 3D Pro
77