background image

31

30

A

lerts

 & A

lArms

: V

isuAl

 

And

 A

udible

 i

ndicAtors

F

unction

/r

esolution

V

isuAl

A

udible

Low Battery
Alarm 

5 audible tones

The 

Recharge Handset

 alarm 

terminates after five seconds 
and goes into auto shutdown. 
Replace the current handset with 
a freshly charged handset. 

NOTE:

 

If the battery voltage is lower than 
required for the X-ray exposure, the 
device will not allow the exposure.

Incomplete 

Exposure
Alarm

A warning tone

This alarm activates if the trigger 
is released before the timed X-ray 
exposure finishes. This condition is 
cleared by pressing any button on 
the user interface panel, by turning 
the power off then on. To avoid this 
error, depress the trigger for the 
duration of the exposure time.

System Failure 

Alarm 

Descending tone

See 

Section 6.2 Troubleshooting 

(6.2.8) or an KaVo authorized 
service center.

Handset 

Failure Alarm 

5 audible tones

See 

Section 6.2 Troubleshooting

 

(6.2.12) or an KaVo authorized 
service center.

The KaVo NOMAD PRO 2 should not be operated if it has been dropped or if 
performance degrades; it should be returned to KaVo for an evaluation.

Содержание NOMAD PRO 2

Страница 1: ...Handheld X ray System for Intraoral Radiographic Imaging Operator Manual TM...

Страница 2: ...nologies LLC its agents or representatives do not accept responsibility for a injury or danger to personnel from X ray exposure b image over under exposure due to poor operating techniques or procedur...

Страница 3: ...18 4 0 Operation 19 4 1 Powering Up 19 4 2 Lock Unlock the X ray 19 4 3 Ensuring the Right Exposure Time is Set 19 4 4 Ready the Device 20 4 5 Initiating and Completing an X ray Exposure 21 4 6 Poweri...

Страница 4: ...Internal and external shielding provide sufficient radiation protection to allow the clinician to remain in the operatory with the patient To make the system as simple as possible for the operator KaV...

Страница 5: ...xposure considering the maturity of the child s physical development If use of the device is considered justified the practitioner must take the appropriate precautions such as use of radiation safety...

Страница 6: ...e product registration card and mail it with proper postage to KaVo today or go on line at kavo com portable x ray machine warranty registration Registering your device will activate your KaVo NOMAD P...

Страница 7: ...he handset charging cradle and AC power supply may become warm to the touch while charging When the handset is first slid onto the charging cradle the first bar will illuminate green for approximately...

Страница 8: ...The KaVo NOMAD PRO 2 provides a high degree of protection from unnecessary radiation However no practical design can provide complete protection nor prevent operators from exposing themselves or othe...

Страница 9: ...imum protection green area from backscatter radiation red area exists when the KaVo NOMAD PRO 2 is positioned near the patient is perpendicular to the operator with the patient s head tilted if needed...

Страница 10: ...limit In response to these regulations KaVo will only ship product that tests below 0 88mGy hr at 1 meter 0 25mGy hr at 1 meter 0 02mGy hr at 5 cm for leakage In order to verify compliance with this l...

Страница 11: ...ints in the room first to establish the vertical significant zone of occupancy and then to establish the horizontal significant zone of occupancy Each exposure was taken at 1 00 seconds in order to si...

Страница 12: ...Connect only items that are specified as part of the system or that are specified as being compatible with the system 2 4 Cleaning Use a non acetone based disinfectant wipe or a cloth to wipe the exte...

Страница 13: ...thorized use In addition it is recommended that the device be locked away when not in use For a further level of security securely store handsets in a separate location Take steps to ensure the KaVo N...

Страница 14: ...n CAUTION X RAYS ATTENTION RAYONS X The positive and negative connections the polarity of a d c power supply Filtration value X ray source assembly Focal Spot Marking indicating that product is compli...

Страница 15: ...he maximum and minimum temperature limits at which the item shall be stored transported or used Kept away from rain and in dry conditions To indicate that the content of the transport package are frag...

Страница 16: ...SA STD C22 2 No 60606 1 Complies with FDA radiation performance standards 21 CFR subchapter J Location Top surface of the NOMAD handheld device Use only with 1 013 2001 1 013 2002 or 1 013 4299 chargi...

Страница 17: ...t the outer cone end Do not attempt to remove this cap or to remove the backscatter shield Attempting to do so will result in damage to your device 3 2 Attaching a Charged Handset With the KaVo NOMAD...

Страница 18: ...for more information Power On Off Attach a fully charged handset to the KaVo NOMAD PRO 2 device Press the Power button and verify that the device correctly powers on and off Increase Decrease Exposure...

Страница 19: ...t button while pressing the Patient button The device shifts into an X rays Locked state and the buttons can be released Repeat the process to unlock the X ray device NOTE If the Select button is held...

Страница 20: ...age before indicating Ready NOTE As a safety precaution if the trigger is held longer than one second or pulled a second time while still in the Enabling X rays state the device will not transition to...

Страница 21: ...ible tone is complete the yellow LED is off and the X ray Complete alert briefly displays The device then returns to the at rest state NOTE An exposure can be terminated for any reason by prematurely...

Страница 22: ...m 90 or perpendicular to the film or sensor To maintain low patient X ray doses and to keep the operator within the protection zone have the head of the patient slightly tilted and or raise or lower t...

Страница 23: ...ray source Time Settings Sensors and Complete Exposures To ensure image quality use correct time settings The KaVo NOMAD PRO 2 comes with pre sets to give you a starting point However these time sett...

Страница 24: ...adults i e the cancer risk per unit dose of ionizing radiation is higher and so unnecessary radiation exposure is of particular concern for pediatric patients Please use caution when configuring the...

Страница 25: ...s the desired menu item s by pressing the Increase and Decrease buttons to scroll up or down When the desired menu item is highlighted press the Select button to access the menu item Within the menu i...

Страница 26: ...r to restore the factory default settings with which the device was shipped These defaults are X rays will be unlocked Audible sound will be set at full volume Trip counter will be reset to zero All f...

Страница 27: ...he collimator cone Carefully align a newly charged handset and while avoiding possible pinch points between the housing and the handset slide it into place A properly oriented handset should snap into...

Страница 28: ...there are any signs of deterioration or if batteries do not retain a charge during normal operation Risk of fire or explosion exists if batteries inside the handset are replaced by unauthorized servic...

Страница 29: ...rts Alarms Visual and Audible Indicators Function Resolution Visual Audible Enabling Alert None This alert is initiated by a single pull and release of the trigger within one second If no further acti...

Страница 30: ...d operation is attempted various alert messages will be displayed For example X rays Locked Simulation Only Simulation Complete Toggle between Locked and Unlocked by pressing and holding down the Sele...

Страница 31: ...vates if the trigger is released before the timed X ray exposure finishes This condition is cleared by pressing any button on the user interface panel by turning the power off then on To avoid this er...

Страница 32: ...d re enable when operator and patient are again properly situated 6 2 3 Enabling X rays alert stays on Double triggering Release the trigger and wait for the Ready alert 6 2 4 Device does not shift to...

Страница 33: ...handset battery can be recharged or is impaired and must be taken out of service A different problem exists if all handsets appear fully charged KaVo NOMAD PRO 2 will require authorized service see S...

Страница 34: ...does not completely charge after 5 hours If the green power indicator is illuminated but the handset fails to charge there may be an issue with compatibility of components Ensure that compatible comp...

Страница 35: ...d by local regulatory authority 2 Review Section 1 1 Indications for Use and product labeling periodically in order to verify understanding of indications for use for the KaVo NOMAD PRO 2 3 A routine...

Страница 36: ...that the trigger moves freely in and out when depressed and released 6 Verify that a single pull and release of the trigger enables the X rays and the device moves into the Ready state 7 Verify that t...

Страница 37: ...constant through the whole exposure time Linearity is calculated per IEC 60601 2 7 50 102 2a Keep the Box This X ray unit may be dangerous to testing technician and any bystanders unless safe test ex...

Страница 38: ...Initial 1 Power Button Schedule Five Year Maintenance Inspection at KaVo 2 Lock out Mode 3 Automatic Shut Off 4 Time Selection Buttons 5 Trigger 6 Enable Ready State 7 Backscatter Shield 8 Exposure Te...

Страница 39: ...Vo or an authorized distributor for replacement and recycling Do not place in municipal waste stream If product return is required contact KaVo for a Return Material Authorization RMA number and shipp...

Страница 40: ...60601 6 62366 60601 2 65 Complies with FDA radiation performance standards 21CFR Subchapter J Internally Powered Type BF MDD 93 42 EEC Annex IX Class IIb IPX specification IPX0 do not operate under w...

Страница 41: ...Exposure time is measured during the entire exposure references to 75 rise fall using the NERO mAx model 8000 X ray meter Collimator Cone Minimum source to skin distance 21cm from focal spot to cone...

Страница 42: ...l spot IEC 60336 1993 0 4mm Maximum anode heat content 4500J Maximum current continuous service 1 5mA x 70kV Maximum anode cooling rate 100W Nominal anode input power 430W Target material Tungsten Min...

Страница 43: ...dimensions are contained entirely within the 6 cm circular exit field diameter This collimator can rotate freely 360 degrees Optional Rectangular Collimator Exit Field Size 3 cm x 4 cm X ray Beam X r...

Страница 44: ...44 Cathode Emission Filament Characteristics Chart Rating Chart Thermal Characteristics Chart...

Страница 45: ...including domestic establishments and those directly connected to public low voltage power supply networks that supply buildings used for domestic purposes The KaVo NOMAD PRO 2 classified as Medical E...

Страница 46: ...PR 11 Group 1 The KaVo NOMAD PRO 2 Intraoral X ray System uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby e...

Страница 47: ...l or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 0 5 1 kV differential mode 0 5 1 2 kV Mains power quality should be that of a typical commercial or hospital envir...

Страница 48: ...in ISM bands E1 3V m Portable and mobile communications equipment should be separated from the KaVo NOMAD PRO 2 Intraoral X ray System by no less than the distances calculated listed below D 3 5 V1 Sq...

Страница 49: ...llular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due...

Страница 50: ...Output Power Watts Separation Distance According to Frequency of Transmitter m Separation m 150kHz to 80MHz D 3 5 V1 Sqrt P Separation m 80 to 800MHz D 3 5 E1 Sqrt P Separation m 800MHz to 2 5GHz D 7...

Страница 51: ...tinue to be warranted for the balance of the original warranty term All warranty claims must be made not later than ten 10 business days following the expiration of the applicable warranty period LIMI...

Страница 52: ...ribex 1 013 3038 EN Rev A 2018 February 1 EC REP MT Promedt Consulting Altenhofstrasse 80 66386 St Ingbert Germany Tel 49 6894 581020 Fax 49 6894 581021 e mail info mt procons com Website www mt proco...

Отзывы: