Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.5 Setting and executing program steps
40 / 76
5.5 Setting and executing program steps
The MASTERsurg LUX Wireless is based on program steps and associated activities
and can be operated intuitively using the graphical user interface.
1
2
Program step 1: Marking
The current program step is shown on the display as number ① and the correspond-
ing activity is shown as symbol ②. Each program step can be assigned to any activity
by selecting the corresponding symbol.
Visualising the activity is an easy means for checking if the activity set on the device is
the same as the current treatment step. Maloperation can thus be largely prevented.
Default values have been set at the factory for the parameters, speeds, torques, trans-
mission ratios and coolant flow rate for every activity according to application. The
parameters can be changed only within a reasonable range for the specific activity. In
the activity, "Free application", all available values can be set. The table below lists the
ranges of values and factory settings.
A treatment sequence can consist of 4 to 10 program steps and/or activities. The
treatment sequence can be designed individually through any arrangement of the
activities. During the sequence, the wireless foot control is used for navigation such
that the device does not have to be touched again during the intervention.
Changed values are saved automatically and are then available for the next use.
5.5.1 Factory settings
The following program steps are pre-set at the factory:
Programme
step
Icon
Activity
Speed [rpm]
Torque [Ncm]
Transmission
ratio
Coolant flow
1
Marking
200 – 2,000
500 (D)
5 – 20
10 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)
2
Pilot drilling
200 – 2,000
500 (D)
5 – 20
10 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)