Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing
23 / 28
7.4 Care products and systems - Servicing
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
CAUTION
Improper service and care.
Risk of injury.
4
Service regularly with suitable agents.
NOTE
KaVo guarantees the proper function of KaVo products only if the care
products listed by KaVo as accessories are used, since these were tested for
proper use on our products.
7.4.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation, but no
later than after 30 minutes of operation.
4
Remove the dental bur from the medical device.
4
Cover the medical device with the KaVo Cleanpac bag, and place it on the
corresponding care adapter.
4
Press the spray key for 1 to 2 seconds.
Servicing the chuck
KaVo recommends servicing the chucking system once weekly.
4
Remove the dental bur from the medical device.
4
Position the tip of the spray nipple in the opening, and apply the spray.
4
Press the spray key for 1 to 2 seconds.
7.4.2 Servicing with KaVo QUATTROcare PLUS
Cleaning and servicing device with expansion pressure for internal cleaning of
inorganic residues and optimum care.
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation, but no
later than after 30 minutes of operation.
4
Remove the dental bur from the medical device.
4
Service the device in the QUATTROcare PLUS.
Also refer to:
Instructions for use KaVo QUATTROcare PLUS
Servicing the chuck
KaVo recommends servicing the chuck system once weekly using the chuck
servicing programme integrated in the device.