7 Reprocessing steps in accordance with ISO 17664 | 7.3 Disinfection
16 / 24
▶ Cover the medical device with the KaVo CLEANpac bag, and place it on the cor‐
responding care adapter. Press the spray button three times for 2 seconds each
time. Remove the medical device from the spray attachment and let the cleaner
work for one minute.
▶ Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also:
2
KaVo CLEANspray / KaVo DRYspray Instructions for Use
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only avail‐
able in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxembourg, Bel‐
gium, Netherlands, United Kingdom, Denmark, Sweden, Finland and Norway.
In other countries interior cleaning can only be carried out with thermodisinfectors in
accordance with EN ISO 15883-1.
7.2.4 Automated internal cleaning
KaVo recommends washer disinfectors in compliance with EN ISO 15883-1, which
are operated using alkaline cleaning agents.
▶ For programme settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the washer disinfector.
▶ In order to prevent negative effects on the medical device, make sure that the in‐
terior and the exterior of the medical device are dry after completion of the cycle,
and then grease it immediately with servicing agents from the KaVo care system.
7.3 Disinfection
WARNING
Hazard due to incomplete disinfection.
Principally, KaVo recommends carrying out an final disinfection of the unpackaged
item if complete disinfection cannot be guaranteed without this measure.
CAUTION
Malfunctioning from using a disinfectant bath or disinfectant containing chlorine.
Defects in the product.
▶ Only disinfect in a thermodisinfector or manually.
CAUTION
Malfunctioning from using the disinfectant bath or chlorine-containing disinfectants.
Defects in the device.
▶ Do not use an ultrasonic bath.
CAUTION
Never use alkaline or chlorine-containing disinfectants.
The saline solution corrodes the metal parts.
▶ Immediately remove all residue.