6 Troubleshooting | 6.2 Troubleshooting
14 / 24
▶ Use the nozzle needle (Mat. no. 0.410.0921) to free the water passage in the
spray nozzles.
6.2.3 Changing the water filter
CAUTION
Hazard from insufficient spray water.
Insufficient spray water can cause the medical device to overheat and damage the
tooth.
▶ Check the filter and exchange if necessary.
▶ Check the spray water channels and clean the spray nozzles with the nozzle nee‐
dle Mat. no. 0.410.0921 if necessary.
▶ Screw out and remove the filter with the wrench (Mat. no. 1.002.0321).
▶ Insert the new filter (Mat. no. 1.002.0271) and screw it in with the wrench.