Troubleshooting
Note
In case of malfunctions, consult the separate instructions for the use and care of
the individual instruments (such as the turbine, motor, camera, Satelec Mini LED,
etc.).
Malfunction
Cause
Remedy
Nothing works.
Main switch is off.
▶ Turn on main switch.
The main fuse is tripped.
▶ Unplug the unit from the mains.
▶ Check the main fuse, and
change it as necessary.
The main fuse is next to the main
switch.
▶ Change the microfuse
(Mat. no. 0.223.2783).
No cold light in the instruments.
Cold light not preselected.
▶ Preselect cold light.
The high-pressure lamp in the in‐
strument is defective.
▶ Replace the high-pressure
lamp.
(See the operating instructions
for the instrument).
No spray in the instruments.
No spray preselected.
▶ Preselect spray.
Close the ring for controlling the
spray on the instruments.
▶ Open the ring for controlling the
spray on the instruments.
The main water valve in the office is
closed.
▶ Open main valve.
The compressor is not turned on.
▶ Turn on the compressor.
Turbine making loud running
noises.
Turbine wheel faulty.
▶ Replace turbine wheels.
Follow the operating instructions
for the turbine.
Spray at the instruments is insuffi‐
cient.
The spray nozzles are dirty/clog‐
ged.
▶ Clean the spray nozzles accord‐
ing to the accompanying instru‐
ment operating instructions.
Leaks in instruments.
O-rings damaged at multiflex or mo‐
tor coupling.
▶ Replace O-rings.
Water in the return air filter.
O-rings damaged at multiflex cou‐
pling.
▶ Replace all O-rings in the multi‐
flex coupling.
Suction hoses not drawing.
Slides on the bodies are closed.
▶ Open slide.
Instructions for use Globus 1078 S
7 Troubleshooting
65/75