Free drainage according to DVGW cer‐
tificate DW-0402 BL 0465
DVGW water block, DVGW water bottle
Water quality
Tap water
Water hardness
1.5 to 2.14 mmol/l ≙ 8.4 to 12 dH
pH
7.2 to 7.8
Customer water filtration
80 µm
Water connection
R 1/2
Water connection above floor
min. 40 mm, max. 75 mm
Water inlet pressure
2.0 to 6.0 bar
Water outlet pressure
5 l/min
Diameter of the drain connection
40 mm
Drainage connection above floor
20 mm
Outflow quantity
max. 5 l/min
Slope of water drain pipe
From the unit, at least 10 mm per meter
Air supply
Air requirements according to DIN EN
7494-2
Dry, oil-free, dirt-free, non-contaminated
Air inlet pressure
5.2 to 7 bar
Air consumption
max. 80 Nl/min.
Customer air filtration
50 µm
Air connection
R 1/2
Air connection above the floor
min. 40 mm, max. 75 mm
Diameter of the suction connection
40 mm
Suction connection above floor
20 mm
Suction vacuum
static at the device inlet: max. 180 mbar,
dynamic: >45 mbar, recommended: 60
mbar
Suction vacuum flow
500 Nl/min.
The values apply to the KaVo measuring set (Mat. no. 0.411.8500)
Operating environment
Floor quality
The quality of the floor must comply with
the load bearing capacity for buildings
DIN 1055 page 3 and the pressure re‐
sistance must correspond to DIN 18560
T 1.
Permissible to a maximum
2 000 m above sea level
Ambient temperature
+10 to +40°C
Relative humidity
30 to 75%
Air pressure
700 to 1,060 hPa