![KaVo Globus 1078 S Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/kavo/globus-1078-s/globus-1078-s_instructions-for-use-manual_3850343014.webp)
CAUTION
Germ formation.
Infections.
▶ Before starting, rinse all the water drain lines without instruments.
▶ Before startup and after the device has not been used for a while (weekends,
holidays, vacations, etc.), rinse or purge the air and water lines.
▶ Thoroughly disinfect.
▶ Actuate the tumbler filler several times.
CAUTION
Third party device connection kit (optional): Hazard of reinfection from standing
water.
Infections.
When a water-using unit is connected to the third-party connection kit, always per‐
form the following tasks on the device:
▶ Before starting, rinse all the water drain lines without instruments (if applicable).
▶ Before startup and after the device has not been used for a while (weekends,
holidays, vacations, etc.), rinse or purge the air and water lines.
▶ Make sure that the water-using unit is resistant to H
2
O
2
since the water is steri‐
lised with OXYGENAL 6 (at a concentration up to 0.02%).
CAUTION
Damage to the instrument hoses from stickers.
Instrument hoses can explode.
▶ Do not affix stickers or adhesive tape.
CAUTION
Long stay in the patient chair.
Decubitus formation.
▶ Take precautions against the formation of decubitus in long treatments.
WARNING
Injury or infection hazard from laid down instruments.
Given the arrangement of the instruments, injury or infections in the hand and un‐
derarm can arise when reaching for the tray holder or operating device. Increased
risk of infection from diseased patients.
▶ Be aware of the arrangement of the instruments when accessing the tray holder
or operating device.
CAUTION
Danger of injury from tipping the treatment unit.
Injury to the patient and user.
▶ Do not support yourself on the swinging arm.
▶ Do not sit on the head or foot end of the patient chair when it is in a horizontal
position.
Instructions for use Globus 1078 S