7 Technical data
40
FOCUS
USE LIMITATION:
External components
WARNING!
Use of accessories, transducers and cables
other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this
device and result in improper operation.
Table 4: Separation distances
Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC
60601-1-2
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The user of the device can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the device as
recommended below, according to the maximum output power of the communications equipment.
Frequency of
Transmitter
Equation
Rated Maximum Output
Power of Transmitter
(watts)
150 kHz to 80 MHz
Separation Distance
(meters)
80 MHz to 800 MHz
Separation Distance
(meters)
800 MHz to 2.7 GHz
Separation Distance
(meters)
0,01
0,20***
0,06***
0,06***
0,1
0,63
0,19***
0,19***
1
2
0,6
0,6
10
6,32
1,90
1,90
100
20
6
6
For transmitters rated at maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTICE!
At 80 MHz, the separation distance for the higher frequency range applies.
NOTICE!
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
WARNING!
*** Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12") to any part of the device, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Содержание FOCUS
Страница 1: ...FOCUS Intraoral X ray User Manual ENGLISH 214760 rev 3 0 805 4936...
Страница 2: ......
Страница 4: ......
Страница 6: ...ii...
Страница 22: ...2 Unit description 16 FOCUS Exposure time control...
Страница 32: ...5 Error messages 26 FOCUS...
Страница 48: ...7 Technical data 42 FOCUS...
Страница 56: ...9 Recommended exposure times 50 FOCUS...