3 Using the unit
FOCUS
19
CAUTION!
Assure that the right image capture unit mode
(Film/Digital) is on.
4.
Press one of the buttons representing the eight
anatomical time settings (tooth buttons). The adjacent
light indicator will illuminate corresponding to the
selection. All other tooth button light indicators will be
off.
5.
The exposure time may be adjusted manually with the
UP and DOWN buttons. The exposure time is based
on the tooth type, patient size, exposure mode (film or
digital), value of kV, film speed and cone length. The
exposure time is shown on the display to two decimal
places. Whenever one of the determining parameters
is changed, the value for the exposure time is
recalculated and the display is updated.
6.
Close the door if a door switch is installed.