12 Details on electromagnetic compatibility | 12.3 Guidelines and manufacturer's declaration - electromagnetic resistance to
jamming
68 / 72
Interference immunity tests IEC 60601 test levels
Compliance level
Electromagnetic environ-
ment - Guidelines
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV contact discharge
± 8 kV atmospheric dis-
charge
± 6 kV contact discharge
± 8 kV atmospheric dis-
charge
Floors should be made of
wood or concrete or be fit-
ted with ceramic tiles. If the
floor is fitted with synthetic
material, the relative hu-
midity must be at least 30
%.
Fast transient electrical in-
terference / bursts accord-
ing to IEC 61000-4-4
± 2 kV for power lines
± 2 kV for power lines
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
Surges according to IEC
61000-4-5
± 1 kV push-pull voltage
(symmetrical)
± 2 kV common mode
voltage
(unsymmetrical)
± 1 kV push-pull voltage
(symmetrical)
± 2 kV common mode
voltage
(unsymmetrical)
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
Voltage interruptions,
short-term interruptions,
and fluctuations of the sup-
ply voltage according to
IEC 61000-4-11
< 5% U
T
for 1/2 period
(> 95% interruption)
40% U
T
for 5 periods
(60 % interruption)
70% U
T
for 25 periods
(30 % interruption)
< 5% U
T
for 5 s
(> 95% interruption)
< 5% U
T
for 1/2 period
(> 95% interruption)
40% U
T
for 5 periods
(60 % interruption)
70% U
T
for 25 periods
(30 % interruption)
< 5% U
T
for 5 s
(> 95% interruption)
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment. If
the user needs the EX-
PERTsurg LUX to work
even if the power supply is
interrupted, we recom-
mend supplying energy to
the EXPERTsurg LUX
from an uninterruptible
power supply or battery.
Magnetic field at a supply
frequency (50/60 Hz) ac-
cording to IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields at the
mains frequency should
correspond to typical val-
ues in a business and
hospital environment.
Note: U
T
is the alternating mains voltage before the application of the test level.
12.3 Guidelines and manufacturer's declaration - electromagnetic
resistance to jamming
The EXPERTsurg LUX is for use in an environment like the one cited below. The EX-
PERTsurg LUX customer or user must ensure that the unit is used in an environment
matching the description.