3.3   Technical Specifications E20 L / E20 C

Drive speed

max. 40,000 rpm

Identification

1 blue ring

Speed transmission

1 : 1

With press-button chuck.

Usable with contra-angle cutters or grinders.

The contra-angle handpiece can be mounted on all INTRAmatic (LUX) mo‐

tors, and motors with a connection in accordance with ISO 3964 / DIN

13940.

 

 

 

 

 

Product description

20

Содержание EXPERTmatic LUX E15 L

Страница 1: ...Instructions for use EXPERTmatic LUX E15 L REF 1 007 5530 EXPERTmatic LUX E20 L REF 1 007 5540 EXPERTmatic E15 C REF 1 007 5531 EXPERTmatic E20 C REF 1 007 5541...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...escription 13 3 1 Purpose Proper use 17 3 2 Technical Specifications E15 L E15 C 19 3 3 Technical Specifications E20 L E20 C 20 3 4 Transportation and storage conditions 21 4 First use 23 4 1 Check th...

Страница 4: ...roubleshooting 37 6 2 1 Troubleshooting Exchanging the O rings on the motor coupling 37 6 2 2 Troubleshooting Cleaning the spray nozzle 39 7 Setup methods according to DIN EN ISO 17664 41 7 1 Preparat...

Страница 5: ...achine disinfection external and internal 50 7 4 Drying 51 7 5 Care products and systems Servicing 52 7 5 1 Care products and systems Servicing Care with KaVo Spray 53 7 5 2 Care products and systems...

Страница 6: ...8 Tools and consumables 63 9 Terms and conditions of warranty 64 Table of contents 4...

Страница 7: ...quality product By following the instructions below you will be able to work smoothly economically and safely Copyright by KaVo Dental GmbH Symbols Refer to the Chapter on Safety Warning symbol Import...

Страница 8: ...quirements of the applicable EC directive Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable Target group This document is intended for dentists and their assistants The se...

Страница 9: ...WARNING Hazard to the care provider and patient In the case of damage irregular running noise excessive vibration un typical warming or when the cutter or grinder is not held firmly Stop working and...

Страница 10: ...dental treatment unit connected must have control equipment for changing the speed and direction of rotation A note is to be included in the documents accompanying the dental treatment unit referring...

Страница 11: ...After treatment place the handpiece properly in the cradle without the cutter or grinder CAUTION Risk of burn injury from hot handpiece head or hot handpiece lid Burn injuries in the mouth may be cau...

Страница 12: ...ical device should be cleaned serviced and stored in a dry location according to instructions before long periods of nonuse CAUTION Hazard from use as a light probe Do not use the medical device as a...

Страница 13: ...ned by KaVo To ensure proper function the medical device must be set up according to the reprocessing methods described in the KaVo Instructions for Use and the care products and care systems describe...

Страница 14: ...Service may only be carried out by KaVo trained repair shops using original KaVo replacement parts Safety 12...

Страница 15: ...3 Product description EXPERTmatic LUX contra angle handpiece E15 L Mat no 1 007 5530 Product description 13...

Страница 16: ...EXPERTmatic LUX contra angle handpiece E20 L Mat no 1 007 5540 Product description 14...

Страница 17: ...EXPERTmatic contra angle handpiece E15 C Mat no 1 007 5531 Product description 15...

Страница 18: ...EXPERTmatic contra angle handpiece E20 C Mat no 1 007 5541 Product description 16...

Страница 19: ...on to the product is impermissible and can be hazardous The medical device is intended for the following applications preparation of cavities caries excavation endodontic applications processing of to...

Страница 20: ...and safety regulations the applicable accident prevention regulations these instructions for use According to these regulations it is the responsibility of the user to only use equipment that is oper...

Страница 21: ...cation 1 green ring Speed transmission 5 4 1 With press button chuck Usable with contra angle cutters or grinders The contra angle handpiece can be mounted on all INTRAmatic LUX mo tors and motors wit...

Страница 22: ...fication 1 blue ring Speed transmission 1 1 With press button chuck Usable with contra angle cutters or grinders The contra angle handpiece can be mounted on all INTRAmatic LUX mo tors and motors with...

Страница 23: ...t has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F Temperature 2...

Страница 24: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 22...

Страница 25: ...First use WARNING Hazard from nonsterile products Infection danger to the care provider and patient Before first use and after each use prepare and sterilise the medical device if needed First use 23...

Страница 26: ...and moist cooling air Contaminated and moist cooling air can cause malfunctions and lead to premature bearing wear Make sure that the supply of cooling air is dry clean and unconta minated according...

Страница 27: ...eck the amount of water CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the dental pulp Adjust the water amount for the spray cooling to a minimum of 50 cm3 min First u...

Страница 28: ...ent spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check the spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 09...

Страница 29: ...l device during treatment A medical device that is not properly locked in place can become discon nected from the motor coupling and fall off Carefully pull on the medical device before each treatment...

Страница 30: ...attach or remove the contra angle handpiece while the drive motor is rotating CAUTION Damage If the footswitch is pressed while you are connecting and removing the medical device you can damage the me...

Страница 31: ...ng with Mehrfunktionsspritze Spray Attach the medical device to the motor coupling and turn it until the guide stud audibly snaps into place Pull on the medical device to make sure that it is securely...

Страница 32: ...d then pulling it along its axis 5 3 Inserting the milling tool or diamond grinder Note Only use carbide cutters or diamond burs that correspond to ISO 1797 1 type 1 are made of steel or tungsten carb...

Страница 33: ...tructions for use and use the cutter or grinder prop erly Only use cutters or grinders that do not deviate from the specified data CAUTION Injury from using worn cutters or grinders Cutters or grinder...

Страница 34: ...from defective chucking system The cutter or grinder could fall out and cause injury Pull on the cutter or grinder to check that the chucking system is okay and the cutter or grinder is securely held...

Страница 35: ...the push button firmly with your thumb and simultaneously insert the milling or grinding tool until it contacts the stop Check that the cutter or grinder is securely attached by pulling on it Operati...

Страница 36: ...er Lacerations and damage to the chucking system Do not touch rotating cutter or grinder Never press the press button while the cutter or grinder is rotating Remove the cutter or grinder from the cont...

Страница 37: ...After the cutter or grinder has stopped rotating press the press button with your thumb and simultaneously pull out the drill bit or bur Operation 35...

Страница 38: ...g or damaged O rings Malfunctions and premature failure Make sure that all O rings are on the coupling and undamaged CAUTION Heating of the product Burns or product damage from overheating Do not use...

Страница 39: ...When the speed drops or is uneven Caring for the medical device An O ring is missing on the motor coupling Replace O ring 6 2 1 Troubleshooting Exchanging the O rings on the motor coupling CAUTION Ha...

Страница 40: ...r coupling may only be lubricated with a cotton ball wetted with KaVo spray Press the O ring between your fingers to form a loop Shove the O ring to the front and remove it Insert new O rings into the...

Страница 41: ...azard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check spray water channels and if necessary clean spray nozzles wit...

Страница 42: ...Clean the water passage in the spray nozzles by using the nozzle needle Mat no 0 410 0921 Troubleshooting 40...

Страница 43: ...es Take suitable personal protective measures Remove all residual cement composite or blood without delay Recondition the medical device as soon as possible after treatment Remove the cutter or grinde...

Страница 44: ...ultrasonic unit Defects in the product Only clean manually or in a thermodisinfector 7 2 1 Cleaning Manual cleaning external Accessories required Tap water 30 C 5 C 86 F 10 F Brush e g medium hard too...

Страница 45: ...with alkaline cleaning agents at a pH value of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD program neodisher mediclean cleaning agent neodisher Z neutralisation agent an...

Страница 46: ...dical device are dry and then lubricate immediately with care agents from the KaVo care system 7 2 3 Cleaning Manual cleaning internal Can only be done with KaVo CLEANspray or KaVo DRYspray Cover the...

Страница 47: ...al interior cleaning are only available in the following countries Germany Austria Switzerland Italy Spain Portugal France Luxem bourg Belgium Netherlands United Kingdom Denmark Sweden Fin land and No...

Страница 48: ...with the VARIO TD program neodisher mediclean cleaning agent neodisher Z neutralisation agent and neodisher mielclear rinsing agent and extends only to the compatibility of materials with respect to K...

Страница 49: ...l device are dry and then lubricate immediately with care agents from the KaVo care system 7 3 Disinfection CAUTION Malfunctioning from using a disinfectant bath or disinfectant containing chlorine De...

Страница 50: ...material compatibility The microbiological efficacy must be ensured by the disinfectant manufac turer Mikrozid AF Liquid made by Sch lke Mayr FD 322 made by D rr CaviCide made by Metrex Consumables r...

Страница 51: ...isinfection internal The efficacy of manual internal disinfection must be demonstrated by the manufacturer of the disinfection agent With KaVo products use only dis infection agents that have been rel...

Страница 52: ...781 G 7881 validation was performed with the VARIO TD program neodisher mediclean cleaning agent neodisher Z neutralisation agent and neodisher mielclear rinsing agent and extends only to the compatib...

Страница 53: ...ith care agents from the KaVo care system 7 4 Drying Manual Drying Blow off the outside and inside with compressed air until water drops are no longer visible Automatic Drying The drying procedure is...

Страница 54: ...in the medical device Risk of injury from sharp or pointed cutters or grinders Remove cutter or grinder CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Pe...

Страница 55: ...7 5 1 Care products and systems Servicing Care with KaVo Spray KaVo recommends servicing the product after each time it is used i e after each automatic cleaning and before each sterilisation Remove c...

Страница 56: ...rvicing the chuck system once a week Remove the cutter or grinder place the spray nipple tip in the opening and spray Carry out the servicing according to the instructions in the section Care with KaV...

Страница 57: ...sterilisation Place the product on the appropriate coupling of the KaVo SPRAYrotor and cover it with a CLEANpac bag Service the product See also Instructions for use KaVo SPRAYrotor 7 5 3 Care produc...

Страница 58: ...automatic cleaning and before each sterilisation Remove cutter or grinder Service the product Care of chucking system KaVo recommends cleaning and servicing the chuck system once a week See also Instr...

Страница 59: ...r or grinder place the spray nipple tip in the opening and spray Subsequently treat with the care products and care systems specified See also Care with KaVo QUATTROcare Setup methods according to DIN...

Страница 60: ...retched The quality and use of the sterilisation packaging must satisfy applicable standards and be suitable for the sterilisation procedure Individually weld the medical device in the sterilised item...

Страница 61: ...in accordance with EN 13060 ISO 17665 1 CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Before each sterilisation cycle service the medical device with Ka...

Страница 62: ...isture Damage to product Immediately remove the product from the steamsteriliserafter the sterilisation cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Setup metho...

Страница 63: ...F 7 4 F Autoclave using the gravitation method at least 10 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F or alternatively at least 60 minutes at 121 C 1 C 4 C 250 F 1 6 F 7 4 F Use according to the manuf...

Страница 64: ...Comply with the expiry date of the sterilised items Setup methods according to DIN EN ISO 17664 62...

Страница 65: ...501 Cellulose pad 100 units 0 411 9862 Cleanpac 10 units 0 411 9691 Nozzle needle 0 410 0921 Material summary Mat no KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spray 2112 A...

Страница 66: ...with a free replacement or repair Other claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only appl...

Страница 67: ...tness of plastic parts All liability is excluded if defects or their consequences originate from ma nipulations or changes to the product made by the customer or a third party that is not authorised b...

Страница 68: ......

Страница 69: ......

Страница 70: ......

Страница 71: ......

Страница 72: ...1 008 2446 kb 20110218 01 en...

Отзывы: