Instructions for use ELECTROmatic M/C and PM/PC
9 Servicing | 9.1 Changing the filter - water inlet
64 / 78
9 Servicing
KaVo recommends to use original KaVo parts
®
for operation and repair only,
since their safety, function, and specific suitability have been tested in extens-
ive tests.
Note
KaVo grants no guarantee of function and KaVo accepts no liability unless ori-
ginal replacement parts and operating materials are used exclusively.
Note
The device must not be serviced or repaired during operation/use.
Servicing work described in the following can be performed by the operator/
user.
9.1 Changing the filter - water inlet
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶
Check and/or replace the filter.
▶
Check the spray water channels and clean the spray nozzles with the nozzle
pin according to need.
KaVo recommends checking the filter every 3 months initially and to change the
filter insert according to need. The scope of delivery includes 2 filter inserts.
After this time, the filter should be regularly checked and replaced according to
need with the interval being appropriate for the conditions at your practice and
the degree of soiling.
▶
Undo the union nut proceeding in the direction of the arrow.
▶
Replace the filter insert if soiling is visible (
Mat. no. 1.007.9736
)
①
.
1
▶
Clean the union nut and the surface of the filter insert.