Instructions for use DEXIS CariVu 3-in-1 by KaVo
7 Processing steps in accordance with ISO 17664 | 7.6 Care products and systems - Servicing
7.5.3 Automated drying
The drying procedure is usually part of the cleaning program of the washer dis-
infector.
Note
Please comply with the instructions for use of the washer disinfector.
▶
In order to prevent impairment of the KaVo medical device, make sure that
the inside and outside of the device is dry after the end of the cycle.
7.6 Care products and systems - Servicing
NOTICE
Improper care.
Malfunction or property damage.
▶
Do not service the medical device with oil or maintenance spray.
7.7 Sterilization
Sterilization in a steam sterilizer (autoclave) in
accordance with ISO 17665-1
Note
The sterile goods package must be large enough for the product so that the
packaging is not stretched.
The quality and use of the sterilization packaging must comply with applica-
ble standards and be suitable for the sterilization procedure!
▶
Seal each medical device individually in a wrap/pouch when sterilizing in a
steam sterilizer (autoclave).
▶
The components/accessories of the DEXIS CariVu 3-in-1 by KaVo are to be
wrapped in an FDA cleared sterilization wrap/pouch when sterilizing in a
steam sterilizer (autoclave).
NOTICE
Contact corrosion due to moisture.
Damage to the product.
▶
Remove the product from the steam sterilization immediately after the
sterilization cycle.
The KaVo medical device has a maximum temperature resistance of up to
138
℃
(280.4 °F).
Sterilization parameters:
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilization processes:
▪ Sterilizer with triple pre-vacuum:
- 3 minutes at 135 °C (275 °F)
Drying time: 16 min.