3.2

   

Technical data

Minimum pressure

2.1 bar (30 psi)

Drive pressure

2.1 to 2.3 bar (30 to 33 psi)

Recommendation

2.1 bar (30 psi)

Air consumption

35 to 40 Nl/min (1.24 to 1.41 cfm)

Idle speed

350,000 to 400,000 rpm

Recommended operating pressure

2 to 3 N (0.45 to 0.68 lbs)

COMPACTtorque 636 turbines can be attached to all hoses with Borden 2-hole

connections.

COMPACTtorque 636 turbines can be attached to all hoses with 4-hole connec‐

tions.

3.3

   

Transportation and storage conditions

CAUTION

It is hazardous to start up the medical device after it has been stored strongly

refrigerated.

This can cause the medical device to malfunction.

▶ Prior to start-up, very cold products must be heated to a temperature of 20°C

to 25°C (68°F to 77°F).

Temperature: -20°C to +70°C (-4°F to +158°F)

Relative humidity: 5% RH to 95% RH absence of condensation

Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)

Protect from moisture

 

 

 

 

 

 

Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830
3 Product description | 3.2 Technical data

6/22

Содержание COMPACTtorque 636

Страница 1: ...Instructions for use COMPACTtorque 636 0 553 0810 COMPACTtorque 636C 0 553 0830 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...7 2 1 Manually cleaning the outside 13 7 2 2 Automated external cleaning 13 7 2 3 Manually cleaning the interior 14 7 2 4 Automated internal cleaning 14 7 3 Disinfection 14 7 3 1 Manually disinfectin...

Страница 4: ...n for users and service technicians Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable CE mark European Community A product bearing this mark meets the requirements of the...

Страница 5: ...ardous situation that can cause damage to property or mild to moderate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maxim...

Страница 6: ...e should be cleaned serviced and stored in a dry location according to instructions before long periods of nonuse Note For safety reasons we recommend that the tool holder system be checked an nually...

Страница 7: ...national statutory regulations Proper use According to these regulations this medical device may only be used for the de scribed application by a knowledgeable user The following must be observed the...

Страница 8: ...with 4 hole connec tions 3 3 Transportation and storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated This can cause the medical devi...

Страница 9: ...and moist cooling air Contaminated and moist cooling air can cause malfunctions and lead to prema ture bearing wear Make sure that the supply of cooling air is dry clean and uncontaminated ac cording...

Страница 10: ...bearing wear Ensure that the cooling air is dry clean and uncontaminated in accordance with EN ISO 7494 2 A minimum drive pressure of 2 1 bar 30 psi is required for operating the medical device The ai...

Страница 11: ...hen the seal is damaged it can cause malfunctions or premature failure Check if the medical device is sealed and undamaged Instructions for use COMPACTtorque 636 0 553 0810 COMPACTtorque 636C 0 553 08...

Страница 12: ...t do not deviate from the specified data CAUTION Injury from using worn cutters or grinders Cutters or grinders can fall out during treatment and injure the patient Never use cutters or grinders with...

Страница 13: ...nder is rotating If you press the pushbutton when the cutter or grinder is rotating it can damage the chucking system and cause injury Never touch soft tissue with the head or tip since it may be hot...

Страница 14: ...o overheat and damage the tooth Check spray water channels and if necessary clean spray nozzles with the nozzle needle Mat no 0 410 0921 Use the nozzle needle Mat no 0 410 0921 to free the water passa...

Страница 15: ...side Accessories required Tap water 30 C 5 C 86 F 10 F Brush e g medium hard toothbrush Brush off under flowing tap water 7 2 2 Automated external cleaning KaVo recommends thermodisinfectors in accord...

Страница 16: ...modisinfec tors in accordance with EN ISO 15883 1 7 2 4 Automated internal cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agent...

Страница 17: ...lubricate the KaVo medical device im mediately with care agents from the KaVo care system 7 3 3 Mechanically disinfecting the exterior and interior KaVo recommends thermodisinfectors in accordance wit...

Страница 18: ...ng and care Reduced product life Perform proper care regularly Note KaVo only guarantees that its products will function properly when the care prod ucts used are those listed as accessories as they w...

Страница 19: ...Unscrew the COMPACTtorque turbine from the hose Cover the product with the Cleanpac bag Insert the tip of the spray nipple into the large tube and press the spray but ton See also Instructions for us...

Страница 20: ...2104A Remove the cutter or grinder place the spray nipple tip in the opening and spray Subsequently treat with the care products and care systems specified See also Servicing with KaVo QUATTROcare 210...

Страница 21: ...front door and start theservice procedure Option 2 After three minutes with no service procedure running the device auto matically switches back to normal service mode See also Care with KaVo QUATTRO...

Страница 22: ...nutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Autoclave using the gravity method at least 10 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F or alternatively at least 90 minutes at 121 C 1 C 4 C 250 F 1 6 F 7 4...

Страница 23: ...10 0921 MULTIflex spray head nozzle 0 411 9921 Material summary Mat no Adaptor INTRAmatic CLEANspray and DRYspray 1 007 1776 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spra...

Страница 24: ...nce non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible...

Страница 25: ......

Страница 26: ...0 589 2232 kb 20130408 02 en...

Отзывы: