Instructions for use MASTERtorque LUX M9000 L - 1.008.7900 MASTERtorque LUX M9000 LS - 1.008.5400
5 Operation | 5.1 Attaching the medical device, M9000 L
5 Operation
CAUTION
Heat transmission during the preparation of abutments.
Thermal damage to the jawbone.
▶
During the preparation of abutments, make sure that the preparation times
are short and that there is sufficient cooling.
Note
At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.
5.1 Attaching the medical device, M9000 L
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked can release from the MULTIflex
coupling during treatment.
▶
Before each use, check if the medical device is securely locked onto the
MULTIflex coupling.
NOTICE
Inaccurate coupling can destroy the high-pressure lamp or the LED of the
coupling or reduce its service life.
▶
Make sure that the turbine is accurately coupled and firmly seated on the
coupling.
▶
Precisely attach the medical device to the MULTIflex coupling and push it to
the rear until the coupling audibly locks in the medical device.
▶
Pull on the medical device to make sure that it is securely affixed to the
coupling.
5.2 Attaching the medical device, M9000 LS
WARNING
Release of the medical device during treatment.
A medical device that is not properly locked can release from the Sirona quick
coupling during treatment.
▶
Before each use, check if the medical device is securely locked onto the
Sirona quick coupling.
NOTICE
Inaccurate coupling can destroy the high-pressure lamp or the LED of the
coupling or reduce its service life.
▶
Make sure that the turbine is accurately coupled and firmly seated on the
coupling.
▶
Precisely attach the medical device to the Sirona quick coupling and push is
to the rear until the coupling audibly locks in the medical device.
17 / 34
Содержание 1.008.5400
Страница 1: ...Instructions for use MASTERtorque LUX M9000 L 1 008 7900 MASTERtorque LUX M9000 LS 1 008 5400...
Страница 32: ......
Страница 33: ......
Страница 34: ...1 009 8239 bd 20200415 06 en...