Instructions for use MASTERtorque LUX M9000 L - 1.008.7900 MASTERtorque LUX M9000 LS - 1.008.5400
4 Start up and shut down | 4.3 Checking the pressures, M9000 L
4.3 Checking the pressures, M9000 L
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion
Premature wear
▶
Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
▶
Insert the test manometer (
Mat. no. 0.411.8731
) between the coupling
and the medical device and check the following pressures:
ð
Drive pressure, drive pressure recommended, return air pressure, spray
water pressure, and spray air pressure.
See also:
2
3.2 Technical Specifications M9000 L, Page 12
4.4 Checking the pressures, M9000 LS
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion
Premature wear
▶
Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
▶
Install the Sirona test manometer test star between the hose and coupling.
Then plug the handpiece onto the Sirona quick coupling and check the fol-
lowing pressures:
ð
Drive pressure, recommended drive pressure, return air pressure, spray
water pressure, and spray air pressure.
See also:
2
3.3 Technical Specifications M9000 LS, Page 12