Instructions for use INTRA Motorhandstück PROPHYwiz 181 P - REF 1.003.2278
6 Reprocessing steps in accordance with ISO 17664 | 6.5 Packaging
6.4.4 Servicing with KaVo QUATTROcare
Note
QUATTROcare 2104 / 2104 A is no longer included in the current de-
livery programme.
Follow-up product:
▶
QUATTROcare PLUS 2124 A
Cleaning and servicing device with expansion pressure for internal cleaning of
inorganic residues and optimum care.
(no validated cleaning of the interior according to German Robert Koch Institute
(RKI) requirements)
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Take the head off.
▶
Unscrew the motor handpiece from the hose.
▶
Service the product in the QUATTROcare.
See also:
2
Instructions for use KaVo QUATTROcare 2104 / 2104A / 2124A
Servicing the clamping chuck
KaVo recommends servicing the chucking system once weekly.
▶
Remove the head and spray into the opening using the tip of the spray
nipple.
▶
Press the spray key for 1 to 2 seconds.
6.5 Packaging
Note
The sterile goods package must be large enough for the handpiece so that
the packaging is not stretched.
The quality and use of the packaging of the items to be sterilised must satisfy
the applicable standards and be appropriate for the sterilisation process!
▶
The medical device must be packed before sterilisation.