Instructions for use INTRA Motorhandstück PROPHYwiz 181 P - REF 1.003.2278
6 Reprocessing steps in accordance with ISO 17664 | 6.4 Care products and systems - Servicing
16 / 22
See also:
2
Instructions for use KaVo QUATTROcare PLUS
Servicing the clamping chuck
KaVo recommends servicing the chuck system once a week using the chuck
servicing program integrated in the device.
Note
Handpieces must be taken off the service couplings before the chuck service
can be started and performed.
▶
Take the head off.
▶
Close the front door and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecut-
ively.
ð
The device is in chuck service mode.
▶
Remove the service coupling of the chuck from the side hatch of the QUAT-
TROcare PLUS and attach it to coupling service point four, on the far right.
A MULTIflex adaptor must be mounted there.
▶
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
▶
Press the button marked with the chuck service symbol.
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start theservice procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
6.4.3 Servicing with KaVo SPRAYrotor
Note
KaVo SPRAYrotor is no longer included in the current delivery pro-
gramme.
Follow-up product:
▶
QUATTROcare PLUS 2124 A
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Take the head off.
▶
Unscrew the motor handpiece from the hose.
▶
Place matching care insert on the appropriate coupling on the KaVo
SPRAYrotor and cover the product with the Cleanpac bag.
▶
Apply spray into the smaller of the two tubes for one second.
See also:
2
Instructions for use KaVo SPRAYrotor