Instructions for use INTRA Motorhandstück PROPHYwiz 181 P - REF 1.003.2278
6 Reprocessing steps in accordance with ISO 17664 | 6.1 Preparation at the site of use
6 Reprocessing steps in accordance with ISO 17664
6.1 Preparation at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶
Take suitable personal protective measures.
▶
Reprocess the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to reprocessing.
▶
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
▶
Remove the prophylaxis head from the medical device.
▶
Remove all residual cement, composite or blood immediately.
▶
Do not place in solutions or similar substance.
6.2 Manual Reprocessing
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
For manual cleaning only.
6.2.1 Manual cleaning - external
Accessories required:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush, e.g. medium-hard toothbrush
▶
Brush off under running tap water.
6.2.2 Manual internal cleaning
Not applicable.
6.2.3 Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
13 / 22