Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912
7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing
▶
Wipe the motor body and handpiece coupling with a dry cloth and remove
any residual oil.
▶
Wipe the inside and outside of the motor sleeve with a water-soaked cotton
cloth and then dry it with a soft cotton cloth.
Note
Cleaning and servicing of the site of coupling.
To maintain uninterrupted operation, it is recommended to clean the grooves
for the O-rings semi-annually. For this purpose, take all 3 O-rings off the
coupling peg and remove all adhering particles and lime scale from the
grooves using a wet, lint-free cotton cloth.
7.2.2 Manual internal cleaning
Not applicable.
7.2.3 Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
NOTICE
Malfunctioning from using a disinfectant bath or chloride-containing
disinfectants.
Defects in the device.
▶
Disinfect manually only!
KaVo recommends the following products based on compatibility of the materi-
als. The microbiological efficacy must be ensured by the disinfectant manufac-
turer and proven by an expert opinion.
Approved disinfectants:
▪ CaviWipes and CaviCide made by Metrex
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
Consumables required:
▪ Cloths for wiping the medical device.