background image

7.6

   

Packaging

Note
The sterilisation bag must be large enough for the handpiece so that the bag is

not stretched.

The quality and use of the sterilisation packaging must satisfy applicable stand‐

ards and be suitable for the sterilisation procedure!

▶ Weld each medical device individually in a sterilised item package!

7.7

   

Sterilisation

Sterilisation in a steam steriliser (Autoclave) EN 13060/ISO 17665-1

CAUTION

Premature wear and malfunctions from improper servicing and care.

Reduced product life.

▶ Before each sterilisation cycle, service the medical device with KaVo care

products.

CAUTION

Contact corrosion due to moisture.

Damage to product.

▶ Immediately remove the product from the steam steriliser after the sterilisa‐

tion cycle!

The medical device has a maximum temperature resistance up to 138℃

(280.4°F).

 

KaVo recommends for example

▪ STERIclave B 2200/ 2200P by KaVo

▪ Citomat/ K-series by Getinge

Autoclave with a pre-vacuum for least four minutes at 134°C ± 1°C

(273°F ± 1.8°F)

Drying time: 20 min.

Autoclave with the gravitation method for at least 10 minutes at 134℃ ±1°C ± 1°C

(273°F ± 1.8°F)

Drying time: 30 min.

Autoclave with the gravitation method for at least 60 minutes at 121° ±1°C

(250°F ± 1.8°F)

Drying time: 30 min.

Follow the manufacturer's instructions for use.

 

 

 

 

 

 

Instructions for use GENTLEsilence 8000BS - 1.000.5600
7 Preparation methods according to ISO 17664 | 7.6 Packaging

17/19

Содержание 1.000.5600

Страница 1: ...Instructions for use GENTLEsilence 8000BS 1 000 5600 Always be on the safe side ...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Страница 3: ... rings on the coupling on the supply hose 12 6 2 Cleaning the spray nozzle 12 6 3 Change the water filter 12 7 Preparation methods according to ISO 17664 13 7 1 Preparation at the site of use 13 7 2 Cleaning 13 7 2 1 Manually cleaning the outside 13 7 2 2 Manually cleaning the interior 13 7 2 3 Mechanically cleaning the exterior and interior 14 7 3 Disinfection 14 7 3 1 Manually disinfecting the e...

Страница 4: ...d of 12 months from the date of the invoice subject to the following conditions In case of justified complaints KaVo will honour its warranty with a free replace ment or repair Other claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only apply in the absence of mandatory legal regulations to the con...

Страница 5: ...orised by KaVo Warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a copy of the invoice or note of delivery The dealer purchase date type and serial number must be clearly evident from this docu ment Instructions for use GENTLEsilence 8000BS 1 000 5600 1 User instructions 1 1 Warranty terms and conditions 3 19 ...

Страница 6: ...RNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury 2 2 Purpose Proper use This medical device is Only intended for dental treatment Any other type of use or alteration to the product is impermissible and can be hazardous The medical device is inten ded for t...

Страница 7: ...hould be cleaned serviced and stored in a dry location according to instructions before long periods of nonuse CAUTION Injury or damage due to wear Irregular running noise significant vibration overheating imbalance or insuffi cient grip Stop work and seek service support Note For safety reasons we recommend that the tool holder system be checked an nually after the warranty period expires CAUTION...

Страница 8: ...pair and service KaVo products Technicians at KaVo branches throughout the world Technicians specially trained by KaVo To ensure proper function the medical device must be set up according to the re processing methods described in the KaVo Instructions for Use and the care products and care systems described therein must be used KaVo recommends specifying a service interval at the dental office fo...

Страница 9: ...and storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F Temperature 20 C to 70 C 4 F to 158 F Relative humidity 5 RH to 95 RH absence of condensation Air pressure 700 hPa to 1060 hPa 1...

Страница 10: ...ly of at least 50 cm min for spray cooling 4 2 Check the pressure A drive pressure of at least 2 7 to 2 9 bar 39 to 42 psi is required for operating the medical device 2 8 bar 40 psi is ideal The air consumption is approx 47 to 53 Nl min Install the Sirona test manometer test star between the hose and coupling Attach the Sirona test manometer test star on the Sirona quick coupling Pressure display...

Страница 11: ...d to the cou pling WARNING Release of the medical device during treatment A medical device that is not properly locked can release from the Sirona quick coupling during treatment Before each use check if the medical device is securely locked onto the Si rona quick coupling 5 2 Remove the medical device Hold the coupling tight and pull the medical device off while twisting slightly 5 3 Insert the m...

Страница 12: ...inder all the way Check that the cutter or grinder is seated by pulling on it CAUTION Hazard from defective chucking system The tool can fall out and cause injury Pull on the tool to check if the chucking system is functioning properly and that the tool is firmly clamped Wear gloves or a thimble to check insert or remove the bits to prevent injury and infection 5 4 Removing the milling tool or dia...

Страница 13: ...jury Never touch soft tissue with the head or tip since it may be hot and cause a burn After treatment is over remove the cutter or grinder from the contra angle handpiece since injury and infection may result from putting it away when the cutter or grinder are inserted Instructions for use GENTLEsilence 8000BS 1 000 5600 5 Operation 5 4 Removing the milling tool or diamond grinder 11 19 ...

Страница 14: ...and exchange if necessary Use the nozzle needle Mat no 0 410 0921 to free the water passage in the spray nozzles 6 3 Change the water filter CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage the tooth If the spray water is insufficient check or exchange the water filter Unscrew the sleeve anticlockwise from the inser...

Страница 15: ...al device The medical device must be dry when transported for reconditioning Do not place it in a solution or similar 7 2 Cleaning CAUTION Malfunctions from cleaning in the ultrasonic unit Defects in the product Only clean manually or in a thermodisinfector 7 2 1 Manually cleaning the outside Accessories required Tap water 30 C 5 C 86 F 10 F Brush e g medium hard toothbrush Brush off under flowing...

Страница 16: ...sinfect in a thermodisinfector or manually 7 3 1 Manually disinfecting the exterior KaVo recommends the following products based on material compatibility Mikrozid AF made by Schülke Mayr liquid or cloths FD 322 made by Dürr CaviCide made by Metrex Consumables required Cloths for wiping off the medical device Spray the disinfectant on a cloth then thoroughly wipe down the medical device and leave ...

Страница 17: ...h compressed air until water drops are no longer visible Automatic Drying The drying procedure is normally part of the cleaning program of the thermodisin fector Note Follow the instructions for use of the thermodesinfector compressed air quality according to ISO 7494 2 7 5 Care products and systems Servicing CAUTION Sharp cutters or grinders in the medical device Injury hazard from sharp or point...

Страница 18: ...7 5 2 Care with KaVo QUATTROcare Cleaning and care unit with expansion pressure for effective cleaning and care Remove the cutter or grinder Servicing the product Servicing the clamping chuck KaVo recommends cleaning and servicing the chuck system once a week See also Instructions for use KaVo QUATTROcare 2104 2104A Remove the cutter or grinder place the spray nipple tip in the opening and spray S...

Страница 19: ...UTION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam steriliser after the sterilisa tion cycle The medical device has a maximum temperature resistance up to 138 280 4 F KaVo recommends for example STERIclave B 2200 2200P by KaVo Citomat K series by Getinge Autoclave with a pre vacuum for least four minutes at 134 C 1 C 273 F 1 8 F Drying time 20 m...

Страница 20: ... protected from germs as far as possible and dust in a dry dark cool room Note Comply with the expiry date of the sterilised items Instructions for use GENTLEsilence 8000BS 1 000 5600 7 Preparation methods according to ISO 17664 7 8 Storage 18 19 ...

Страница 21: ... 003 6884 Cleanpac 10 units 0 411 9691 STERIclave bags 0 411 9912 Nozzle needle 0 410 0921 Water filter with nipple 1 000 4823 SIRONA spray head for KAVOspray 1 005 8365 KAVOspray 2112 A 0 411 9640 SIRONA turbine treatment coupling or QUATTROcare 1 000 7156 QUATTROcare plus Spray 2140 1 005 4525 Instructions for use GENTLEsilence 8000BS 1 000 5600 8 Accessories 19 19 ...

Страница 22: ......

Страница 23: ...1 006 2887 kb 20081009 03 en ...

Отзывы: