6
Troubleshooting
6.1
Exchanging the O-rings on the coupling on the supply hose
CAUTION
Hazard from improper care of the O-rings.
Improper care of the O-ring can cause the medical device to partially or com‐
pletely malfunction.
▶ See the instructions for use for the Sirona quick coupling.
6.2
Cleaning the spray nozzle.
CAUTION
Hazard from insufficient spray water.
Overheating of the medical device and damage to the tooth.
▶ Check the spray water channels and clean the spray nozzles with the nozzle
needle Mat. no. 0.410.0931 if necessary.
▶ Check the water filter and exchange if necessary.
▶ Use the nozzle needle (Mat. no. 0.410.0921) to free the water passage in the
spray nozzles.
6.3
Change the water filter
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶ If the spray water is insufficient, check or exchange the water filter.
▶ Unscrew the sleeve ① anticlockwise from the insert ② and pull off.
▶ Pull back the insert ② slightly.
▶ Use tweezers to remove the water filter ③ from the insert ②.
▶ Insert a new water filter.
▶ Pull back the insert ② slightly.
▶ Place the sleeve ① on the insert ②, and screw it tight clockwise.
Instructions for use GENTLEsilence 8000BS - 1.000.5600
6 Troubleshooting | 6.1 Exchanging the O-rings on the coupling on the supply hose
12/19