4 Start up and shut down | 4.1 Mount the MULTIflex (LUX) / MULTIflex LED coupling
4 Start up and shut down
WARNING
Hazard from contaminated products.
Infection hazard for care provider and patient.
▶
Prior to initial commissioning and after each use, reprocess the product and
accessories.
See also:
2
7 Reprocessing methods according to ISO 17664, Page 20
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
▶
Reprocess and sterilise the product and accessories before disposal.
See also:
2
7 Reprocessing methods according to ISO 17664, Page 20
NOTICE
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions.
▶
Make sure that the supply of cooling air is dry, clean and uncontaminated
according to ISO 7494-2.
4.1 Mount the MULTIflex (LUX) / MULTIflex LED
coupling
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked can release from the MULTIflex
coupling during treatment.
▶
Before each use, check if the medical device is securely locked onto the
MULTIflex coupling.
▶
Screw the MULTIflex (LUX) / MULTIflex LED coupling onto the turbine hose
and tighten it with the wrench
Mat. no. 0.411.1563
.
▶
Rotate the spray ring on the MULTIflex coupling (465 series) in order to
regulate the water supply.
4.2 Checking the amount of water
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Thermal damage to the dental pulp.
▶
Adjust the water amount for the spray cooling to a minimum of 50 ml/min!
13 / 28