7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
20 / 28
7 Reprocessing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶
Take suitable personal protective measures.
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
▶
Reprocess the medical device as soon as possible after treatment.
▶
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
▶
Remove the tool from the medical device.
▶
Remove all residual cement, composite or blood immediately.
▶
Do not place in solutions or similar substances.
7.2 Manual reprocessing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
For manual cleaning only.
7.2.1 Manual cleaning - external
Accessories required:
▪ Tap water
▪ Brush, e.g. medium-hard toothbrush
▶
Brush off under flowing tap water.
7.2.2 Manual internal cleaning
Not applicable.