Instructions for use INTRAflex LUX 2 Programm 2307 LN - 0.535.6190, 2320 LN - 0.535.6180, 2325 LN -
0.535.6210
4 Start up and shut down | 4.3 Checking the pressures
4.3 Checking the pressures
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion
Premature wear
▶
Supply dry, clean and uncontaminated compressed air according to ISO
7494-2 only.
▶
Insert the test manometer (
Mat. no. 0.411.8731
) between the coupling
and the medical device and check the following pressures:
ð
Drive pressure, drive pressure recommended, return air pressure, spray
water pressure, and spray air pressure.
Drive pressure
2.5 to 3.0 bar (29 to 44 psi)
Recommended drive pressure
3.0 bar (44 psi)
Return air pressure
< 1 bar (15 psi)
Spray water pressure
1.0 to 2.0 bar (15 to 29 psi)
Spray air pressure
1.5 to 2.5 bar (22 to 36 psi)
4.4 Checking the O-rings
NOTICE
Missing or damaged O-rings.
Malfunction and premature failure.
▶
Make sure that all O-rings are on the coupling and are undamaged.
Number of available O-rings: 5
15 / 28