81
3.
Remove the back panel.
Spring position
adjustment
This explains how to adjust the furnace door springs to a proper tension.
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
1.
Disconnect the instrument from the wall outlet.
2.
Remove the furnace back/top panel (see page 80,
3.
Open the door manually and hold it fully open.
4.
Be careful in the next step, because springs are quite strong.
5.
Adjust the lower position of the springs by re-attaching them to a different
position on the spring adjustment hook. Always adjust left and right springs
symmetrically.
6.
Test the new spring adjustment, as described at page 57,
Furnace door
removal
Removing the furnace door can be necessary for cleaning or maintenance
operations.