68
Mold holder
installation
1.
With the mold holder components at hand, proceed like the mold holder’s
middle section removal, but in reverse order.
2.
However, do not install the front ceramic tube immediately, as it could fall
during the crucible holder installation. Wait until the crucible holder is installed,
then install the mold holder’s front ceramic tube. You may however install the
card-shaped metallic spacers, which set the position of that tube.
3.
Check adjustments as described below.
Adjustment of
holders
To keep your instrument in good working condition, adjustment of the holders
must be checked periodically, and each time a part of the holders is replaced.
IMPORTANT
: Incorrectly adjusted holders can cause pouring issues and
damage the instrument. Please follow each step in order, to ensure proper
adjustment.
Except where noted, all fasteners must be screwed in tightly after adjustment.
Turn the power off before starting this procedure, so that you can manually move
the motors when needed. The holders should be fully assembled on the fluxer
before starting this procedure.
Vertical position of the crucible axles
1.
The desired height of each crucible axle is such that it is vertically centered
within the furnace slot (i.e. with the same distance to the slot, above and
under the axles).
Adjustment