K-3 & K-4 Series Owner’s Manual
Page 37
K-4oem ~ AIRE-FLOAT™ QUICK REFERENCE GUIDE
K-4oem ~ AIRE-FLOAT™
ALTERNATING TRUE LOW AIR LOSS SYSTEM
THIS DOCUMENT IS FOR QUICK REFERENCE ONLY. PLEASE READ THE MANUAL BEFORE OPERATING THE K-4oem SYSTEM.
Note:
Before using the KAP MEDICAL Mattress or Overlay System, please remove current mattress,
mattress replacement systems or overlay systems from the bed.
•
Mattress / Overlay set up.
Make sure the mattress/overlay fits snugly on the bed frame without any gaps on the side.
•
Overlay System: Place the K-4oem overlay directly on an existing 3” to 5” foam mattress. Insert head and
foot end elastic straps around the foam mattress. Loop each long side strap around the foam mattress and
fasten it securely to the short strap on the other side using the buckle. Hang the K-4oem control unit on
the footboard and make sure that the mattress hose assembly is connected securely to the control unit.
•
Mattress Replacement: Unroll the K-4oem mattress replacement and place it directly on the bed frame
and attach it firmly with the straps provided. Hang the K-4oem control unit on the footboard, make sure
that the mattress CPR, bolster and/or air pad plugs are connected properly and check that the mattress
hose assembly is connected securely to the control unit.
•
Plug the hospital grade power cord provided with the unit into a three-pronged hospital socket. The amber
light will illuminate. Press POWER key
(SB)
to green light position and then press MAX Inflate
(W).
The
blower will come on and quickly inflate the mattress.
•
Using the SOFT / FIRM keys
(K)
set patient comfort pressure settings. 7 segment display will show patient
comfort settings from 0 (lowest pressure) to 9 (highest pressure)
.
•
For ALTERNATING PRESSURE
(AP)
therapy, press Mode
(M)
key until the desired A/P time LED
(AP)
lights up.
•
To set STATIC
(T)
mode, press Mode key
(M)
until “STATIC” LED
(T)
lights up.
•
To set ALARM SILENCE, press the ALARM SILENCE key
(AS)
until the indicator light goes on.
•
To set the LOCK OUT Function, press the LOCK OUT key
(LO)
until indicator light goes on.
•
K-4oem is equipped with a Fowler key
(U)
which when pressed in the patient’s fowler position of 35 degrees
or greater will inflate the mattress to higher pressures to eliminate patient bottoming.
•
For
CPR
or quick deflation, disconnect mattress hose connector from the control unit by simply pulling the
quick connector from the control unit in a straight outward direction.
Service: Clean filter periodically to prevent filter clogs.
♦
For more information, please call authorized service center
♦