K-3 & K-4 Series Owner’s Manual
Page 31
CONTROL UNIT CALIBRATION AND SETTINGS
K-3 ELITE & K-3 OEM SERIES
K-3 ELITE
K-3oem
Calibration
(Control Unit must be
in Stand-By mode and
the mattress hose
disconnected)
•
Press and hold “Alarm Silence”, “Soft” & “Firm”
keys until unit beeps. (Approx. 3 seconds).
•
Unit will display calibration instructions.
•
Press and hold “Alarm Silence”, “Mode” & “Lock”
keys until unit beeps. (Approx. 3 seconds). Static
and Pulse LEDs will flash.
•
Press “Select” key complete calibration and end
routine.
Auto Fowler Set-up
(Control Unit must be
in Stand-By mode)
•
Place new Fowler Transmitter (in bed or external)
to the flat position.
•
Press and hold “Patient Setup” & “Mode” keys
until unit beeps. (Approx. 3 seconds).
•
To end routine without new fowler, press “Power”
key.
•
Routine will auto time out in 2 minutes if no
transmitter is detected.
•
Place new Fowler Transmitter (in bed or external)
to the flat position.
•
Press and hold “Fowler” & “Lock” keys until unit
beeps. (Approx. 3 seconds). “L” will be shown on
the 7- segment display.
•
Tilt transmitter/bed to fowler position, unit beeps,
all LEDs light up and unit returns to stand-by mode
•
To end routine without new fowler, press “Fowler”
key.
•
Routine will auto time out in 60 seconds if no
transmitter is detected.
Low Pulse Time
Set-up
(Control Unit must be
in Stand-By mode)
•
Press and Hold “Pulse(N)” and
“Patient set up(H)” keys until unit beeps and
displays Pulse menu (Approx. 3 seconds)
•
Press the (L) “+/-” keys to adjust low pulse
time (from 4 to 90 seconds)
•
Routing will time out in 20 seconds or press the
Power key to end routine.
•
Set-up not available on K-3oem model. (default
time: 30 seconds pulse cycle and 4 seconds low
pulse time)
K-4 ELITE & K-4 OEM SERIES
K-4 ELITE
K-4oem
Calibration
(Control Unit must be
in Stand-By mode and
the mattress hose
disconnected)
•
Press and hold “Alarm Silence”, “Soft” & “Firm”
keys until unit beeps. (Approx. 3 seconds).
•
Unit will display calibration instructions.
•
Press and hold “Alarm Silence”, “Mode” & “Lock”
keys until unit beeps. (Approx. 3 seconds). Static
LED will flash for about 3-5 seconds.
•
Once Static LED stops flashing, press “Mode” key
to complete calibration and end routine.
Auto Fowler Set-up
(Control Unit must be
in Stand-By mode)
•
Place new Fowler Transmitter (in bed or external)
to the flat position.
•
Press and hold “Patient Setup” & “Mode” keys
until unit beeps. (Approx. 3 seconds).
•
To end routine without new fowler, press “Power”
key.
•
Routine will auto time out in 2 minutes if no
transmitter is detected.
•
Place new Fowler Transmitter (in bed or external)
to the flat position.
•
Press and hold “Fowler” & “Max” keys until unit
beeps. (Approx. 3 seconds). “L” will be shown on
the 7-segment display.
•
Tilt transmitter/bed to fowler position, unit beeps,
all LEDs light up and unit returns to stand-by mode
•
To end routine without new fowler, press “Fowler”
key.
•
Routine will auto time out in 60 seconds if no
transmitter is detected.
Set A/P Low Value
K-4 Elite
(Control Unit
Must be Powered on)
K-4oem
(Control Unit must be
in Stand-By mode)
•
Press “Alternating Pressure (AP)” key to activate
AP therapy.
•
Press “Mode” key, A/P settings will be displayed
on the screen.
•
Use the “+” and “-“ keys to adjust the Low AP
percentage value (10%-75%)
•
Unit will time out and setting will be set
automatically within 15 seconds.
•
Press and hold “Mode” & “Fowler” keys until unit
beeps. (Approx. 3 seconds).
•
7- segment will display
“0” (0% mmHg of high comfort pressure setting) or
“5“ (50% mmHg of high comfort pressure setting)
•
To end routine and accept setting, press “Power”
key.
•
Routine will auto time out in 30 seconds if no key
is pressed.