K-3 & K-4 Series Owner’s Manual
Page 20
DEFAULT 10 MIN ALTERNATING PRESSURE
•
Press (D) key to set 10 min auto default (AP) cycle. (K-4 Elite model only)
PULSE (N)
(WAVE)
•
(K-3 Elite and K-3oem models) This mode is selected by pressing the (S) Select key until the pulse LED lights up
(in K-3oem model) or by pressing the (N) Pulse key (K-3 Elite model).
•
During Pulse mode, the pressure in each air cell will be decreased by 20% for 5 seconds and then increase to the prior
pressure setting for 30 seconds. This cycle will continue every 30 seconds as long as pulsation is enabled. For model
K-3, use the (L) “+/-” key to select the desired pulsation interval time.
•
(K-3 Elite model only) To adjust the low pulse time interval for Pulse therapy, follow the next steps:
•
(Perform in Stand-By mode) Press and Hold “Pulse(N)” and “Patient set up(H)” keys until unit beeps and displays
Pulse menu (Approx. 3 seconds).
•
Use the (L) “+/-” keys to adjust low pulse time (from 4 to 90 seconds).
•
Routine will time out in 20 seconds from the last key pressed, or press the Power key to end routine.
MASSAGE
•
(K-4 Elite model only) To set Massage mode, press Mode key (M)
until the display reads “MASSAGE”.
•
In this mode the patient receives rapid alternating pressure (rapid high pressures and low pressures).
MULTI-THERAPY
•
(K-4 Elite and K-3 Elite models only) To set Multi therapy mode, press Mode key (M)
until the display reads “MULTI”.
Use the TIME ADJUST keys (L) “+/-” to set the desired “Multi” therapy time.
•
Multi-therapy is the combination of Static, A/P, Pulse and Massage therapies in K-4 Elite model and combination of Static
and Pulse therapies in K-3 Elite model.
MODE KEY (M)
•
The Mode key is used to cycle between therapy modes in K-4 Elite, K-3 Elite and K-4 oem models. In addition to cycling
between therapy modes, in K-4 Elite and K-3 Elite models the Mode key is used to configure the Patient Set-Up
function. In K-4 Elite model only, the Mode key is also used to access the Alternating Pressure settings.
PAN (Y)
•
(K-3 only) This mode is selected by pressing the (N) Pan key. Pan mode enables for quick deflation of center section of
mattress for bed pan operation.