60
Trouble Shooting
Output check (1) - Print out
Problem
Possible cause(s) / Solution(s)
Reference
The printer doesn’t
operate.
The printer is not connected properly.
Confirm the printer connection.
Page 34
The Baud Rate might not be set properly.
Check both the instrument and printer
settings.
Page 46
Make sure to use the optional printer (Model
DPU-S245) and printer cable (Model
6000-31).
Page 34
Confirm the connection procedure.
Connect the printer to the printer port,
then turn on the instrument and printer.
Page 34
Cannot printout the
display
Display is not frozen.
1. Press
to hold the display.
2. Press
to print out.
Page 34
Want to cancel printing
You can’t cancel print while it is operating.
Page 34
Output check (2) - Digital output
Problem
Possible cause(s) / Solution(s)
Reference
PC doesn’t load the
measuring data.
Confirm the USB connection, which is
different from printer connection.
Page 39
The Baud Rate might not be set properly.
Check both the instrument and PC
settings.
Page 34
Confirm the communication command (ex.
Hyper Terminal command).
Page 39
Output check (3) - Analog output
Problem
Possible cause(s) / Solution(s)
Reference
No data is generated.
Check the connection and confirm the
polarity.
Page 43
The HOLD function might be active.
Press
to cancel the HOLD function.
Page 43
Data is tiered.
Data is outputted every one second.
Page 43
The output data is not
correct.
Confirm the setting for analog output.
Page 43
Confirm the output range.
Page 43
The load impedance might not be optimum.
Set the load impedance larger than 5k
Ω.
Page 43