![Kanomax Climomaster S-6531-0E Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/kanomax/climomaster-s-6531-0e/climomaster-s-6531-0e_user-manual_652703061.webp)
59
§ 15
Trouble Shooting
Check battery
Problem
Possible cause(s) / Solution(s)
Reference
The instrument will not
turn on
Battery is inserted in wrong polarity.
Turn off the power and insert the battery
correctly.
Page 11 &
13
The display is not visible
Display contrast is not set up properly.
Adjust the contrast volume switch.
Page 50
The battery level indicator
flashes on the display
The battery is low.
Turn off the instrument and replace the
batteries.
Page 11 &
13
Initial operational check
Problem
Possible cause(s) / Solution(s)
Reference
The display is not visible
The display contrast is not set up properly.
Adjust the contrast dial.
Page 50
[NO PROBE] is displayed
The probe is not connected.
Turn off the instrument and connect a
probe.
Page 12
Can I change the
measurement unit?
Yes, you can change measurement units by
following the steps on page 46.
Page 46
Normal operational check
Problem
Possible cause(s) / Solution(s)
Reference
[**.*] is displayed as
measured value.
It indicates the measurement is outside the
measuring range.
The instrument must be used in the range
stated in the specifications.
Page 52
[- - - - ] is displayed as air
velocity measurement.
The probe or the probe cable is not connected
properly.
Check the connection between the probe
and the probe cable.
Page 12
The probe or the probe cable may be
damaged.
Contact your local Kanomax
representative for repair or replacement.
Page 62
The instrument is not
reading the correct air
speed.
The probe might be not facing into the wind.
Make sure the wind direction indicator is
facing into the wind.
Page 14
The instrument reads a
higher temperature with
no wind.
CLIMOMASTER can’t measure the correct
temperature with no wind (less than 0.1m/s
of wind). *Model 6533 and 6543 can
measure less than 0.1 m/s.
Page 14
The response time is slow Set the correct time constant (TC).
Page 20
[ - - - - ] is displayed as
the air volumetric flow
measurement.
The duct information (shape and size) is not
stored.
Register duct information.
Page 16
actoolsupply.com
actoolsupply.com
Содержание Climomaster S-6531-0E
Страница 1: ...actoolsupply com ...