Normal Measurement
NORMAL MODE
Setting the Time Constant
In NORMAL Mode, you can change the
Time Constant (TC) by pressing
.
You can select the Time Constant from 1, 5 or 10seconds.
TC1: Displays the instantaneous value at every 1 second
TC5: Displays moving average for 5 seconds
TC10: Displays moving average for 10 seconds
*The Time Constant is only effective in Normal Mode. When you first turn
on the instrument, it is effective only for the AIR VELOCITY and the AIR
FLOW measurements in Normal Mode. If you need to apply the time
constant to the AIR TEMPERATURE, HUMIDITY and PRESSURE
measurements in Normal Mode, refer to the next section “Changing
Time Constant Application”.
*** What is Time Constant? ***
Time Constant determines the time span of the moving average. When you set the Time
Constant to a larger (longer) value, the readings will become stable. When you select a smaller
(shorter) value, the readings will become more responsive and sensitive to changes in air
velocity. This function is not available in Calculation or Flow Rate Mode.
MODE
How To Take In Measured Data
EXPLANATION
TC1
Displays the
instantaneous value at
every 1 sec.
TC5
Updates every second.
The value displayed is
the average of the past 5
seconds.
TC10
Updates every second.
The value displayed is
the average of the past
10 seconds.
'12/04/12
a
15:40:23 ~
TC
C
1
0.87f
25.3e
64.1g
Time Constant
Data for 1 sec
Average of 5 sec
Average of 10 sec
■
Humidity is only
available on
MODEL6531/6533.
■
Air temperature is
only available on
MODEL6531/6541
/6542/6533/6543/
6561.
sec.
(Measuring Time)
sec.
(Measuring Time)
sec.
(Measuring Time)
■
Pressure is only
available on
MODEL6501-B0/-C0.
actoolsupply.com
actoolsupply.com