219 Route 206 Phone: 1-800-247-8887
PO Box 372
Fax: (973) 786-7586
Andover, NJ 07821 USA [email protected]
6824
™
User Manual, PN 10283
Kanomax USA, Inc
Rev 1.0, 30-July-2021
www.kanomax-usa.com
Copyright © 2021, Kanomax USA, Inc.
Page 16 of 28
TM
APPENDIX D – ANALOG OUTPUTS (IF EQUIPPED)
If the instrument is equipped with the analog output option, there will be a five-
pin connector on the bottom of the instrument. Also, an analog output cable will
be included with the instrument. This cable will have a five-pin connector on
one end and four tinned wires on the other end.
The instrument will output a Voltage between 0 and 5 Volts that corresponds to
the Air Velocity and Temperature measured by the instrument. The output
range, pin assignments, and wire colors are given in the table below. Also
shown is a block diagram of the analog output circuit.
Corresponding
Wire
Analog
Output
Measurement
Equation* to convert from
Color
Pin
#
Function
Voltage
Range
Range
Volts (
V
) to Measurement
Value
BLK
1
Ground
--- N/A ---
--- N/A ---
--- N/A ---
GRN
2
Air Velocity
0 to 5
Volts
0 to 10,000
FPM
Air Velocity = 2000×
V
GRY
3
--- Not Used ---
--- N/A ---
--- N/A ---
--- N/A ---
WHT
4
Temperature
0 to 5
Volts
-20°C to 80°C
Temperature [°C] = 20×
V
- 20
*To convert from Volts to Air Velocity in Feet per Minute, multiply by 2000.
Ź)
or example, an analog output of 2.375 Volts on the green wire means that
the instrument is measuring an air velocity of 4750 feet per minute (FPM).
Ź9ROWV
×
2000 = 4750 FPM
*To convert from Volts to Temperature in degrees Celsius (°C), multiply by 20
and subtract 20.
Ź)RUH[DPSOHDQDQDORJRXWSXWRI9ROWVRQWKHZKLWHZLUHPHDQVWKDW
the instrument is measuring a temperature of 49.16 °C.
Ź9ROWV
×
20 – 20 = 49.16 °C